Software
5.5.3
Manual Start of the Pump Test
It is also possible to check the function of the spare pump manually. To start the pump
test manually:
1. Select the
OFF
mode in the menu bar.
2. Select
Tools
-
Options
-
Pump Control
-
Pump Test
.
3. Ensure that the checkbox „Enable Automatic Pump Test“ is checked.
4. Press the „Start pump test now“ button (
1).
5. The pump test will proceed as described in the automatic test. A status bar and text
output will show the status of the pump test cycle.
During the test the P1 field will display the P1 pressure and chart the values over
time. Likewise the helium flow will be displayed in the He Flow field and the values
over time will be charted. In both cases the x-axis will be adjusted automatically
Figure 5.20 Tools - Options - Control Unit - Pump Test
i
If the automatic pump test is disabled, a faulty backup pump will not be recognized. Thus
when the main pump fails there will be no backup up pump which could lead to instabili-
ties of the magnet. This will also trigger an alarm cycle.
1
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......