66
ID: 182463115
Planning details
If the system is installed within an HF shielded room, suitable connecters, extensions and HF
filtered feed through elements can be ordered under the
Bruker
part number T11976
.
6.10
Cold water supply
Part of the heat dissipated from the MR system is released via a heat exchanger to a cold water
supply system.
Heat dissipated to the cold water system
Heat dissipated to the cold water system
Table 6.20: Heat dissipated to the cold water system
1
Extraction tube for anesthetic gas (inner diameter of the tube=25 mm) and connection
to the white plastic duct 90° bend with inner diameter of 110 x 54 mm.
2
Set for anesthetic gas extraction system
Standard configuration
MRI CryoProbe™
option
Heat dissipated into the water
~ 11 kW
Additional ~ 8.5 kW
Supply temperature
8-12 °C (*)
8 – 24 °C
Maximum pressure (forward flow)
15 bar
7 bar
Minimum pressure difference (forward
flow)
5 bar (*)
3.2 bar
Minimum flow
45 l/min (*)
7 l/min
Recommended particle filter
500 µm
500 µm
Содержание BioSpec 117 AVANCE III
Страница 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Страница 6: ...Contents 6 ID 182463115...
Страница 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Страница 72: ...72 ID 182463115 Planning details...
Страница 82: ...82 ID 182463115 Delivery and transport...
Страница 84: ...84 ID 182463115 Moving or dismantling the MR system...
Страница 92: ...Figures 92 ID 182463115...
Страница 94: ...Tables 94 ID 182463115...
Страница 101: ...ID 182463115 107 Lastpage...
Страница 102: ...Bruker Corporation info bruker com www bruker com...