Planning details
Care should be taken in both cases to ensure that the supporting and contact surfaces are ab-
solutely plane. The contract surfaces must be sealed prior to assembly of the filter plate with HF
sealing tape.
Assembly of filter plates with an installation frame (right) and directly screwed to the cage (left)
Figure 6.5: Assembly of filter plates with an installation frame (right) and directly screwed to the cage (left)
1
sheet copper of the Faraday cage
2
external insulation of the shielding box
3
Allen screw with spring ring and washer
4
HF sealing tape
5
filter plate
6
inner lining and inner frame of the filter plate opening
7
aluminum frame on which to place the filter plate
8
press nut
H
Height of the filter plate
C
Width of the supporting surface
B3
Distance to bore hole