Installation
Figure 1.5. BCU05 connections
4. Connect the nitrogen (or dry air) input from the BVT.
5. Set the air flow rate to approximately 500 l/h at the BVT. Check that the sample
spinning is not affected. If yes, decrease the air flow rate.
Note: Please install the unit only on a hard floor.
6. If the BCU05 is controlled by a BVT3000/3300, connect the BVT plug to the
connector called «BCU05» located on the BVT's front panel.
7. In stand alone mode, don't connect the BVT plug and close the switch located
behind the rear grill.
Figure 1.6. Rear grill
filter
heat exchanger
dried air or N2 in
mains
BVT connector
switch
serial number plate
rear grill
Содержание BCU05
Страница 1: ...BRUKER Version VTU Technical Manual BCU05 006 ...
Страница 4: ...4 41 BRUKER BCU05 Version 006 Contents ...
Страница 6: ...6 41 BRUKER BCU05 Version 006 Index ...
Страница 18: ...18 41 BRUKER BCU05 Version 006 BCU05 HR applications ...
Страница 22: ...22 41 BRUKER BCU05 Version 006 BCU05 HR MAS applications Figure 2 3 BCU05 connections W4M123712A ...
Страница 24: ...24 41 BRUKER BCU05 Version 006 BCU05 HR to BCU05 HR MAS ...
Страница 27: ...BCU05 Version 006 BRUKER 27 41 5 Schematics 5 ...
Страница 30: ...30 41 BRUKER BCU05 Version 006 Figure 5 3 BCU in operation W4M123717A support stand ...
Страница 33: ...BCU05 Version 006 BRUKER 33 41 Figure 5 6 BCU05 HR MAS Connection diagram ...
Страница 35: ...BCU05 Version 006 BRUKER 35 41 Figure 5 8 BCU05 in operation ...
Страница 36: ...36 41 BRUKER BCU05 Version 006 Schematics ...
Страница 38: ...38 41 BRUKER BCU05 Version 006 Figures ...
Страница 40: ...40 41 BRUKER BCU05 Version 006 Tables ...
Страница 41: ...BCU05 Version 006 BRUKER 41 41 Lastpage ...