background image

  %

  

  

  %

44

  &"

 #"! !$! #

(98/12/10)

Address

Byte Format

3 2 1 0

Read/Write

Destination

1F0A0014

MODE0

W

- - - x b

1F0A0014

MODE0

R

- - - x b

1F0A0018

MODE1

W

- - - x b

1F0A0018

MODE1

R

- - - x b

1F0A001C

MODE2

W

- - - x b

1F0A001C

MODE2

R

- - - x b

4. 2. 9.  Real Time Clock and NVRAM

Notes

   The MK48T02 of SGS–Thomson is installed to implement these two

functions by one device.

   It is a 2K–Byte–Static–RAM with its eight upmost cells being reserved

as hold register for the real time clock information.

   The lower half of the 2 kbyte is normally protected and can only be writ-

ten by means of the special PAL device ”CU09Nx01 ” of the service
staff. So far, this range contains only the ”ethernet physical address”.

   Besides the timing registers, the upper half of NVRAM contains all the

other boot parameters read with ”printenv” and changed with ”setenv” in
the monitor program.

Device Codes

Address

Destination

Read/Write

Byte Format

3 2 1 0

1F00000x

to

1F006FFx

NVRAM range, lower half

R

x x x b

1F00700x

to

1F007F7x

NVRAM range, upper half

R/W

x x x b

1F007F8x

Control Register

R/W

x x x b

1F007F9x

Seconds

R/W

x x x b

1F007FAx

Minutes

R/W

x x x b

1F007FBx

Hour

R/W

x x x b

1F007FCx

Day

R/W

x x x b

1F007FDx

Date

R/W

x x x b

Содержание AQS

Страница 1: ...BRUKER Version Communication Control Unit CCU 9 10 AQS AQX Technical Manual 001...

Страница 2: ...the manual leading to coincidental damage whether during installation or operation of the instrument Un authorised reproduction of manual contents without written permission from the publishers or tra...

Страница 3: ...AQX CCU 9_4x00 AQS CCU 10_4x00...

Страница 4: ...s 19 2 11 Power Requirements 22 3 Service Information 23 3 1 Operating Conditions 23 3 2 Network connection 23 3 3 Software Implementation 24 3 4 Problems of Point to Point Connection to the O2 25 3 5...

Страница 5: ...igure 1 CU09 Controls at Front Panel 6 Figure 2 CU10 Controls at Front Panel 7 Figure 3 CU09 Attaching SubD connectors to the RS232 485 interfaces 10 Figure 4 CU 10 Attaching SubD connectors to the RS...

Страница 6: ...ssories 8 Table 6 CU09 Configuration of Serial Interfaces 10 Table 7 CU10 Configuration of Serial Interfaces 11 Table 8 Parameters to be set 15 Table 9 DRAM configuration 16 Table 10 RS232 Signal assi...

Страница 7: ...are connected inside to the same signals but have crossed pin assignments The upper socket signed FETH HUB has to be used if CU09 10 will be connected to an HUB This is the normal way with twisted pa...

Страница 8: ...up depending on the interface to which the spect is connected to 1 4 Problems of Point to Point Connection to the O2 Both link partners of a Fast Ethernet Connection exchange their abilities 10 or 100...

Страница 9: ...ns Type on O2 as superuser vi etc config netif options This file contains the 2 parameter entries of if1name ec0 and if2name ec1 Swap the values of both parameters to if1name ec1 if2name ec0 Type wq i...

Страница 10: ...memory to IO data path is 32 bit wide and runs at 25 MHz AQX Software Constrains CCU Versions Part No Layout No EC Level AspectStation SGI Computers 2 2 Former AQS CCU Versions The AQS_CCU is a CPU b...

Страница 11: ...RS485 channels using Z85C230 labeled as tty10 and tty20 connectable through one 8 pin Mini D console and two 50 pin con nectors tty01 tty09 at the front edge real time clock MK48T02 with 2 kbyte non v...

Страница 12: ...Wire BNC connector Reset Button RS232 console Mini D 8 Activity LED s TTY01 TTY05 TTY10 TTY06 TTY09 TTY20 10 100 BASE T Fast Ethernet Status LED s to Hub s Fast Ethernet connector RJ 45 to Host VME ac...

Страница 13: ...00 BASE T Fast Ethernet Status LED s to Hub s Fast Ethernet connector RJ 45 to Host VME access Memory access 10Base2 Ethernet activity Collision occurence Good Link status Receive activity Transmit ac...

Страница 14: ...U09 Fast Ethernet UTP CAT5 RJ45 Cable 5m 83025 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 10m 83026 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 0 5m 67980 CU09 Fast Ethernet UTP CAT5 RJ45 Cable 3m 67981 CU09 F...

Страница 15: ...ne or two cables of the following type SCSI Cable P N 73104 Connecting to RS232 RS485 of CU10 The RS485 Channels tty10 and tty20 are only avalable at the AQS Backplane Connec tor X3 Connecting the RS2...

Страница 16: ...6 tty07 tty08 tty09 tty01 tty02 tty03 tty04 tty05 not used not used CAN tty06 tty07 tty08 tty09 Connections made with SCSI Cable P N 73104 Thin wire Ethernet BNC Connector RJ45 Connector to other Host...

Страница 17: ...tty08 tty09 tty01 tty02 tty03 tty04 tty05 not used not used CAN tty06 tty07 tty08 tty09 Connections made with SCSI Cable P N 73104 Thin wire Ethernet BNC Connector RJ45 Connector to other Hosts St5 RJ...

Страница 18: ...c tion CU09 Terminal console cable 3m P N HZ04112 This cable was originally designed to connect a plotter to the Indy Figure 5 Console Cable MINI D 8 pin to D_SUB 9 pin Pin Signal 1 DTR 2 CTS 3 TxD 4...

Страница 19: ...flect the activities of the Ethernet DMA the CPU access to DRAM and the activity of the CPU on VME bus The intention is that after power up or reset a basic test procedure changes the LED value at suc...

Страница 20: ...le set of diagnostic tests and sets the NVRAM parameters to the following default values netaddr 0x149 236 99 99 netmask 0xffffff00 lbaut 9600 rbaut 9600 bootfile bfs usr diskless clients spect root u...

Страница 21: ...or clears the memory after reset and boots without doing any test d The boot procedure has to be started by typing u or auto the memory will not be cleared and no tests will be carried out eth_addr_ch...

Страница 22: ...sition A disables the SQE test mode e g if there is a Repeater connected 2 9 3 Installing DRAM It is possible to equip the CCU with 16 32 or 64 Mbyte using 2 of 8 Mbyte DRAM modules P N 65173 16 Mbyte...

Страница 23: ...W5 W4 W3 Test jumper W6 DRAM DRAM W2 W1 W7 High Low Firmware PROM s U75 U76...

Страница 24: ...W1 High Low Firmware PROM s W7 U75 U76 W5 W4 W3 Test jumper W6 DRAM DRAM W2...

Страница 25: ...ons CU09 10 INDY O2 Terminal Mini DIN 8 Mini DIN 8 D SUB 9 D SUB 25 Shield housing housing housing 1 DTR 1 5 CTS 2 6 7 20 TxD 3 5 2 3 GND 4 4 5 7 RxD 5 3 3 2 RTS 6 2 8 6 DCD 7 4 GND 8 CU09 ST8 ST9 tty...

Страница 26: ...CU10 ST8 ST9 tty01 tty09 RS232 Connector CU10 ST2 tty10 tty20 RS485 JTAG and BBIS Connector...

Страница 27: ...standard twisted pair cable has a straight through pin to pin connection from plug to plug The star topology of twisted pair Ethernet connecting several Hosts via one Hub requires a different signal...

Страница 28: ...oltages Part No 5 V 12 V 12 V 5 V analog J3 C8 5 V analog J3 C1 C5 CCU CU09 10 H2570 H9503 5 3 A 0 3 A 0 1 A 0 0 CU09 supplies fused current at the following connectors ST8 9 RS485 12 Volt 0 2 Ampere...

Страница 29: ...can be carried out as on CU09 with the exception of the RS485 interface that is now available on the AQS backplane The BNC cable of the Thin Wire Ethernet 10Base2 should be left unconnected 10Base2 is...

Страница 30: ...ne of the two link partners is down by software or power The Fast Ethernet cable can be pulled out and plugged in under power and activity Autonegotiation and link checking starts on both partners wit...

Страница 31: ...reases the performance by at least 10 and leads to collisions Furthermore the point to point connection with O2 causes errors during the boot se quence This problem can be solved by connecting the O2...

Страница 32: ...r diskless clients spect root unix This call leads to booting over 10Base2 The CU09 10 monitor will boot over Fast Ether net if fe is inserted instead of la or as default if no name is set boot f bfs...

Страница 33: ...OUI Device Group Series Production 0000ad 0000 09 3 7 Introduction Status of the AQX CU09 CCU 3 7 1 Modifications of the introduced layout There are 25 prototypes of CU09 using layout H3P2160C with s...

Страница 34: ...on 970501 included XWIN NMR 2 0 re lease 3 8 Introduction Status of the AQS CCU10 3 8 1 Modifications of the introduced layout The new AQS CCU CU10 board contains the same functions as the previous CU...

Страница 35: ...5 CCU Connector ST7 ST8 ST9 ST9 ST2 ST2 AQX Ext U Connector N A ST2 ST 6 ST7 ST 8 ST9 ST10 N A N A Pin out of the X3 connector at the AQS backplane Signal Signal 2 TxD _1 1 TxD _1 4 GND 3 GND 6 RxD _1...

Страница 36: ...to be virgin Starting with Firmware Version 199990408 the Power on Diagnostics will be first carried out at the second boot procedure If magic RISCPROM the Firmware will boot normally but if the para...

Страница 37: ...th_addr_ch1 0x0000ad03xx06 eth_addr_ch2 0x0000adxxxx09 version 5 40 magic RISCPROM model CCU R4600 vendor BRUKER rootname 0 use_bootparams 1 keyswtch 0 keyboard AT 3 8 3 2 Firmware CU09 10 needs the f...

Страница 38: ...rate 15 to 255 Reserved and zero The serial boot prom is listed in the PAL part list It is named CU09EA00 and labeled EA or 50MHz 3 9 AQX CU09 CCU History of Modifications EC No Date Part Number Descr...

Страница 39: ...can access only the Memory 3 Slave devices which are accessible only by the CPU The communication between these Sections is provided by Interconnect Interface This is a CCU s Internal Interface resolv...

Страница 40: ...Data reg Triport bus Exchanger registered P L M rocessor Bus Bus emory Bus Ae L Address Bus L Data Bus P Bus Pipeline r Memory Write multil reg 4 x 64 bit Memory 64 64 Bit Wide 32 9 De 32 M_D Mma P_D...

Страница 41: ...e referred to as word 32 bit operands with the valid bytes indi cated by b at Byte format see below Byte 0 means the byte connected to data lines D0 to D7 Byte halfword 2 byte word 4 byte and block 4...

Страница 42: ...This RAM has to be en abled by setting jumper W3 and is then located in the lowest address space starting with address 0x0 The installed memory volume must be properly selected by setting of jumper W...

Страница 43: ...e Pagemode is not recommended because of its impact to the on board ethernet operation Single word writes and reads use a single read write register A pipeline structure is made for multiple writes an...

Страница 44: ...eserved R W bbbb 1F08003C CFIT Interrupt R W bbbb undefined 1F080040 CFDA Driver area R W bbbb undefined Control and Status Register Device Codes The CSRs registers are accessibly only after loading a...

Страница 45: ...and Fiber Optic cables The chip interfaces the PMD sublayer through DP82223 Twisted Pair Transceiver to the MAC layer 21140A through a Media Independent Interface MII The 32 word wide MII status and...

Страница 46: ...to 0001 05h ANLPAR 5 Does not reflect Link Partner 10BASE T Abil ity after parallel detection Reflects Link Partner 10BASE T Ability after parallel detection 05h ANLPAR 7 Does not reflect Link Partner...

Страница 47: ...6 21 1 6 5 762 26 6 21 5 1 19h 11 5 48 55 46 55 46 6 4 5 4 0 4 6 0 276 Differences between DP83840 and 83840A in full duplex mode 00h 9 9 01h 9 9 02h 9 9 03h 9 9 8 5 21 2 04h 9 9 05h 9 9 1 461 4 06h...

Страница 48: ...F022xxx Channel A Control Register cons R W x x x b 1F023xxx Channel A Data Register cons R W x x x b 1F024xxx Channel D Control Register tty03 R W x x x b 1F025xxx Channel D Data Register tty03 R W x...

Страница 49: ...to 1F0A0000 1F0A3FFF address region JTAG submodul is compatible with IEEE 1149 1 JTAG Standard Test Access Port and Boundary Scan Architecture Supported National s SCAN Ease Embedded Application Softw...

Страница 50: ...rmally protected and can only be writ ten by means of the special PAL device CU09Nx01 of the service staff So far this range contains only the ethernet physical address Besides the timing registers th...

Страница 51: ...s not available if the respective digit equals to zero All configuration register outputs are provided in their negated form The STATUS register accommodates the temperature and VME bus time out statu...

Страница 52: ...ssigned and hardwired to the floating point in terrupt The router distributes all onboard and vme bus interrupt sources to level 0 to 4 Source interrupts intended to a certain level are ored to produc...

Страница 53: ...d their location in the corresponding interrupt regis ter are as follows Interrupt Level Location INT_0 INT_1 INT_2 INT_3 INT_4 in Reg IRQ1 IRQ4 IRQ6 0 IRQ7 D0 TMP_INT IRQ3 IRQ5 0 F_INT D1 res TIM_INT...

Страница 54: ...implemented W x x x b 4 2 12 CIS access The information EEPROM of the Component Identification System is connected to the CCU con trolled CIS channel The EEPROM of all CCU s have one and the same uni...

Отзывы: