Technical Data MS 500‘89 Ascend
9.5
Nitrogen Level Sensor
The Nitrogen Level Sensor is inserted in the recommended nitrogen turret. 5 lights
display the nitrogen level (
).
Nitrogen Level Sensor
Material No.
Value
Unit
Level Sensor Type
Z122404
12/1126/860 D325
Diameter
12
mm
Overall length
1126
mm
Active length
860
mm
Table 9.7: Nitrogen level sensor
Figure 9.4:
Nitrogen level sensor
Nitrogen Level at:
more than 90 %
70 % – 90 %
50 % – 70 %
30 % – 50 %
10 % – 30 %
Schedule Refill
less than 10 %
Refill immediately
Содержание 500'89 Ascend Aeon
Страница 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Страница 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Страница 8: ...8 ZTKS0172 Z31964 04 Contact ...
Страница 28: ...28 ZTKS0172 Z31964 04 Safety ...
Страница 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Страница 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Страница 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Страница 81: ...81 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend 9 21 Current Lead Figure 9 12 Current lead 55 pin ...
Страница 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Страница 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Страница 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Страница 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Страница 98: ...98 ZTKS0172 Z31964 04 Index ...
Страница 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...