48
ZTKS0269 / Z33108 / 02
Troubleshooting
Continued from page before
6.2.5
During Operation of the Magnet Stand
In case of doubt contact Bruker Service and refer to the manual of the Magnet Stand.
Continued on next page
Indicator
Possible reason
Solution
By
Magnet system does
not reach
specification.
Magnetic material
inside RT bore tube.
Carefully clean the RT bore
tube.
Bruker Service
Large ferromagnetic
parts near the
magnet system.
1.
Keep the maximum
possible distance
between the magnet
system and
ferromagnetic parts.
2.
Repeat shimming.
Bruker Service
Indicator
Possible reason
Solution
By
The NMR spectrum
shows massive
disturbances.
Pneumatic controller
is in DOWN position.
Switch pneumatic controller to
UP position.
Approved
Customer
Personnel
Magnet system has
direct mechanical
contact with the floor
via accessories or
cables.
Identify and eliminate contact
point. Arrange cables in loose
S- or U-shapes.
Approved
Customer
Personnel
If the problem is still not
solved, contact
Bruker Service.
Approved
Customer
Personnel
Magnet system has
physical contact to
the magnet stand.
Check leveling; adjust if
necessary.
Bruker Service
Piston of the isolator
is not centric or
touches its casing.
Align magnet stand.
Bruker Service
T-safety bracket
touches the pillar.
Align magnet stand.
Bruker Service
Floor vibrations in
vertical direction.
Replace elastomeric isolators
with air damped isolators.
Bruker Service
Floor vibrations in
horizontal and
vertical direction.
Replace air damped isolators
with air piston isolators.
Bruker Service
Содержание 300'154 Ascend DNP
Страница 1: ...Version Innovation with Integrity 300 154 Ascend DNP 02 User Manual NMR...
Страница 6: ...6 ZTKS0269 Z33108 02 Table of Contents...
Страница 8: ...8 ZTKS0269 Z33108 02 Contact...
Страница 32: ...32 ZTKS0269 Z33108 02 Transportation...
Страница 33: ...33 ZTKS0269 Z33108 02 4 Assembling 4 1 Safety Approved Persons Bruker Service only...
Страница 34: ...34 ZTKS0269 Z33108 02 Assembling...
Страница 38: ...38 ZTKS0269 Z33108 02 Operation...
Страница 54: ...54 ZTKS0269 Z33108 02 Troubleshooting...
Страница 57: ...57 ZTKS0269 Z33108 02 8 Disassembling 8 1 Safety Approved Persons Bruker Service only...
Страница 58: ...58 ZTKS0269 Z33108 02 Disassembling...
Страница 66: ...66 ZTKS0269 Z33108 02 Technical Data MS 300 154 Ascend DNP 9 6 Helium Level Graph Figure 9 4 Helium Level Graph...
Страница 74: ...74 ZTKS0269 Z33108 02 Technical Data MS 300 154 Ascend DNP 9 13 Current Lead Figure 9 10 Current Lead 55 pin...
Страница 94: ...94 ZTKS0269 Z33108 02 Appendix...
Страница 96: ...96 ZTKS0269 Z33108 02 Appendix...
Страница 98: ...98 ZTKS0269 Z33108 02 Appendix...
Страница 100: ...100 ZTKS0269 Z33108 02 Index...
Страница 102: ...Bruker Corporation info bruker com www bruker com Order No Z33108...