12 (47)
BRUKER BIOSPIN
Operating & Service Manual Version 001
Installation
Environment Requirements
3.1.4
This amplifier is built for inside use only on a maximum elevation of 2000m above
sea level (6600 feet).
No specific cooling or ventilation is required.
Be sure that the amplifier has enough area around it so that the free airs flow into
and out of the amplifier is not obstructed.
It should, however, be in an environment which conforms to the 5°C - 45°C (41°F
- 113°F) thermal specifications, a 80% maximum relative humidity of air and a
contamination level of two (means a normal non-conductive contamination,
temporary conductivity due to condensation is possible).
Installation Requirements
3.2
No special precautions are necessary. Mount the equipment in an area which is
relatively free of vibration, and has sufficient room for cable connections.
The amplifier has a class II installation category.
Bench Operation
3.2.1
The unit can be placed onto a secure flat surface.
Power Requirements
3.3
The BLAH2000-I E C Amplifier 300MHz has a build-in switched power supply.
The mains line connector on the rear panel is a CEI 10A.
One Phase Line requirements :
AC input voltage :
208-230VAC
Input current max :
5.5A
Frequency :
50/60Hz
System Check
3.4
Before applying power for the first time the following items should be checked:
•
The AC input voltage 208-230 VAC ± 10% range must be compatible with the
power supply.
•
AC line is connected.
•
CAN interface is connected.
•
Power forward and reflected are connected.
•
The BLAH2000-I E C Amplifier 300MHz has a nominal input level of 0dBm.
Ensure that the system drivers are operating at these levels.
•
Output RF loads are connected.
Содержание BLAH2000-I E C
Страница 6: ...6 47 BRUKER BIOSPIN Operating Service Manual Version 001 General Information ...
Страница 10: ...10 47 BRUKER BIOSPIN Operating Service Manual Version 001 Safety ...
Страница 14: ...14 47 BRUKER BIOSPIN Operating Service Manual Version 001 Installation ...
Страница 20: ...20 47 BRUKER BIOSPIN Operating Service Manual Version 001 Operation ...
Страница 44: ...44 47 BRUKER BIOSPIN Operating Service Manual Version 001 Figures ...
Страница 46: ...46 47 BRUKER BIOSPIN Operating Service Manual Version 001 Tables ...
Страница 47: ...Operating Service Manual Version 001 BRUKER BIOSPIN 47 47 End of Document ...