PULSE Multi-analyzer System Type 3560-B/C/D/E –
Installation and IDAe Hardware
84
6.13 Generator, 4/2-ch. Input/Output Module Type 3109
Stop
BNC connector allowing connection to an emergency stop control,
stopping the generators immediately. A simple switch that shorts the
signal line to ground will enable this.
LEDs
Measuring:
Indicates a measurement in progress or the downloading
of software to the module.
Overload:
Common to all four input channels, this LED indicates the
status of the channels (activated: green; overload: red). See section 7.8.
Outputs 1 and 2:
These are green if the output is active.
Outputs 1 and 2
BNC connectors (see Fig.7.5 on page 90) for output of a high-quality
generator signal over the frequency range to 25.6 kHz. The maximum
output voltage is 5 V
rms
delivered in one output range through a 24-bit
D/A converter. Output may be grounded or floating, controlled by
PULSE software.
Waveforms are determined by PULSE software and include:
•
Sine – fixed or swept (burst or continuous)
•
Dual sine – fixed, swept or combination
•
Random (burst or continuous)
•
Pseudo-random
•
Periodic Random
•
User-defined waveform (import from WAV file)
Inputs 1 to 4
For input of signals in seven ranges from 7.071 mV to 7.071 V
peak
in 10 dB steps.
Note:
The LEMO and BNT/BNC connectors of each input channel are connected in parallel.
You should therefore only connect one transducer to each pair of connectors at any one
time.
Input 1:
BNT connector (see Fig.7.3 on page 89) for direct connection of a voltage signal
or for connection of CCLD transducers and 7-pin LEMO connector (see Fig.7.6 on page 90)
for use with Brüel & Kjær microphones. The BNT connector also provides power for a tacho
probe.
Содержание PULSE 3560-B
Страница 2: ......
Страница 8: ......
Страница 11: ...Part I Installation...
Страница 12: ......
Страница 44: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 36...
Страница 53: ...Part II Hardware...
Страница 54: ......
Страница 110: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 102...
Страница 140: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 132...
Страница 142: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 134...
Страница 143: ...Part III Appendices...
Страница 144: ......
Страница 148: ...PULSE Multi analyzer System Type 3560 B C D E Installation and IDAe Hardware 140...
Страница 161: ......