Calibration
21
Connect Cables with
PHOTON
+
and Start Software
Cable Connection for Calibration
Connect the BNC cables to all input channels and the OUTPUT channel
at one end. Connect all the other ends together with BNC T ports.
Connect this cable harness to a voltmeter that is certified to the accuracy
standard described above (0.25%).
Power up the
PHOTON
+
and wait for at least 10 minutes prior to
continuing. Record the environmental parameters such as pressure,
humidity, etc. if necessary.
Start the Calibration Software
Start the Calibration software by clicking the Calibration icon.
Select
New
under the
File
menu:
The user is asked to enter the following information regarding this
calibration: