Chapter 2 — Checking Procedure
Checking Procedure
2238 Mediator Integrating Sound Level Meter
Service Manual
BI 5043 –13
13
Table 2.1 The relation between the Display reading and the DC and AC Output
2.3.3
Frequency Weighting
Action:
●
Set the generator to 1000 Hz and adjust the output so that Type 2238 reads
128.0 dB
in SPL for Linear (LLFP) mode.
●
You must readjust the output to 128.0 dB at 1000 Hz when you change weighting
to A - Weighted (LAFP) and again when you change to C - Weighted (LCFP). To
change weightings, press
, then
to
Weightings
and
to select. Use
to
navigate to the
Freq. Wgt.
line, and
to the next desired weighting. Use
to
OK
and
to close and return to the active display.
Table 2.2 Linear (LLFP)
Table 2.3 A-Weighted (LAFP)
Input
dBµVRMS
SPL Reading
on
Display
DC Output
mV
Actual
Reading
AC Output
mV
Actual
Reading
approx. 126
REF. 130.0
4000 ± 50
1000 ± 50
Generator Freq.
Hz
dB SPL value on the
Display
Actual
Reading
Ref 1000
128.0
128.0
10
127.7
±
1.0
19950
126.0
±
1.0
Generator Freq.
Hz
dB SPL value
on the Display
Actual
Reading
Ref 1000
128.0
128.0
15.85
71.3 +1.0/ – 3.0
63.1
101.8
±
0.5
3981
129.0
±
0.4
19950
118.7 + 1.0/ – 3.0
Содержание Mediator 2238
Страница 1: ...BI 5043 13 Technical Documentation 2238 Mediator Integrating Sound Level Meter Service Manual...
Страница 2: ......
Страница 7: ...2238 Mediator Integrating Sound Level Meter Service Manual BI 5043 13 1 Chapter 1 Introduction...
Страница 14: ...2238 Mediator Integrating Sound Level Meter Service Manual Br el Kj r 8...
Страница 15: ...2238 Mediator Integrating Sound Level Meter Service Manual BI 5043 13 9 Chapter 2 Checking Procedure...
Страница 27: ...2238 Mediator Integrating Sound Level Meter Service Manual BI 5043 13 21 Chapter 3 Mechanical Parts...