APPENDIX A
Tables
35
Table A.44
Weatherproof Microphone Unit Type 4184-A is not specified together with Outdoor Microphone Kit
UA-1404
Table A.45
Weatherproof Microphone Unit Type 4184-A is not specified together with Outdoor Microphone Kit
UA-1404
Table A.46
Weatherproof Microphone Unit Type 4184-A is not specified together with Outdoor Microphone Kit
UA-1404
Table A.47
Weatherproof Microphone Unit Type 4184-A is not specified together with Outdoor Microphone Kit
UA-1404
Table A.48
Weatherproof Microphone Unit Type 4184-A is not specified together with Outdoor Microphone Kit
UA-1404
Periodic Testing of Acoustical Frequency Responses
This section gives the adjustment data that must be applied to sound levels displayed in response to the
sound pressure produced by Multifunction Acoustic Calibrator Type 4226, in order to obtain the
equivalent sound levels that would be displayed under reference environmental conditions in response to
plane progressive sinusoidal sound waves incident from the reference direction. See Table A.49 to view
the data.
The uncertainty of the corrections in Table A.49 includes a component for inter-sample variability of
Weatherproof Microphone Unit Type 4184-A. Due to the construction of the microphone, the inter-sample
variability cannot be separated from the adjustment data. The uncertainty is allowed for in the adjustment
of Weatherproof Microphone Unit Type 4184-A.
NOTE:
that the data in Table A.49 are only valid at the frequencies stated in the table. The data values can
not
be interpolated for other frequency values.
Table A.49
Acoustical test with Multifunction Acoustic Calibrator Type 4226. Adjustment data that must be applied to
the readings of the analyzer in order to obtain equivalent sound levels that would be displayed in response to plane
progressive sinusoidal sound waves incident from the reference direction
Nominal
Exact
Correction Data
Expanded
Correction Data
Expanded
Correction Data
Expanded
Frequency
Frequency
For 4184-A 0
Uncertainty
For 4184-A 90
Uncertainty
for Diffuse Field
Uncertainty
(6 digits)
(Top)
(Side)
Hz
Hz
dB
dB
dB
dB
dB
31.5
31.6228
0.00
0.05
0.00
0.05
0.00
0.05
63
63.0957
0.00
0.05
0.00
0.05
0.00
0.05
125
125.893
0.00
0.05
0.00
0.05
0.00
0.05
250
251.189
0.00
0.05
0.00
0.05
0.00
0.05
500
501.187
0.10
0.16
–0.09
0.16
–0.01
0.16
1000
1000.00
0.19
0.14
0.07
0.14
0.12
0.15
2000
1995.26
0.36
0.19
–0.06
0.19
0.22
0.21
4000
3981.07
0.75
0.32
–0.71
0.32
–0.40
0.36
8000
7943.28
0.52
0.53
–1.81
0.53
–1.32
0.56
12500
12589.3
1.46
0.76
–1.85
0.76
–1.07
0.78
16000
15848.9
0.80
1.30
–2.38
1.30
–1.65
1.32
Table A.50
No data are given for Acoustical Test with Electrostatic Actuator UA-0033
Содержание 4184-A
Страница 2: ......
Страница 6: ......
Страница 42: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 36...
Страница 48: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 42...
Страница 50: ...Weatherproof Microphone Unit Type 4184 A Supplement to Instruction Manual BE 1712 44...
Страница 51: ......