Chapter 4 – Measuring
Starting a New Measurement
BB 0908–13
4 – 9
Precision Integrating Sound Level Meter Type 2236
User Manual
that resetting will erase all previous measurement re-
sults from the buffer. If this occurs, press
〈
OK
〉
to confirm
that you want to reset the sound level meter.
The results in the sound level meter’s display buffer are
erased and the elapsed time is set to zero. The overload
hold is reset so that the sound level meter indicates that
there have not been any overloads since the last reset.
Note: To erase all results in the Log, Memory and display
buffer, press
and
〈
Data
〉
.
4.
If you want to measure according to IEC standards (i.e.
free field), simply point the sound level meter towards
the sound source.
If you want to measure according to ANSI standards (i.e.
diffuse field), fit the supplied Random Incidence Corrector
DZ9566 on the microphone. The direction of the sound
level meter is unimportant. If, however, the sound field
is free, measure with the sound level meter at an angle
of between 70 and 80˚ to the sound source.
5.
Press
.
The sound level meter starts measuring with the selected
set-up.
Note:
●
When mounting the sound level meter on a tripod, posi-
tion the tripod so that one of its legs points in the same
direction as the sound level meter. This will reduce the
risk of damaging the microphone if the tripod is acciden-
tally knocked over.
●
See section 1.3 for practical hints and information about
measuring according to standards.
▼