Instructions for Use

Overview of CoolBox

 2XT System

CoolBox 2XT is designed for maintaining sample temperatures below 4°C on the bench top with-
out the use of wet ice or electricity. Patent-pending dual-phase conductive XT Cooling Core or XT 
Freezing Core provides the cooling source when a thermo-conductive CoolRack

®

 or CoolSink

®

 tube 

or plate module is placed on top. The core and sample module in combination ensure uniform 
well-to-well temperature throughout the cooling period regardless of sample position. XT Cooling 
Core can maintain sample temperature from 0.5 to 4°C for over 16 hours; XT Freezing Core can 
maintain frozen samples for 10 hours; using dry ice in the base provides an ultra-cold temperature 
(-78°C) for snap-freezing samples in tubes or plates. For a list of CoolRack and CoolSink modules 
that are compatible with the CoolBox 2XT, see back page.

Assembly of CoolBox

 2XT

The CoolBox 2XT system consists of a closed-cell, cross-linked polyethylene foam base, collar, lid 
and a reusable XT Cooling Core. A CoolBox 2XT also includes a CoolRack

®

 or CoolSink

®

 tube or 

plate module. When placed on top of the XT Cooling or Freezing Core, the modules and samples 
rapidly equilibrate to the temperature of the XT Core. Samples can be placed in the CoolRack 
or CoolSink module either before or after the module is placed in CoolBox XT. When used with 
CoolSink plate modules the magnetized collar is not required and can be removed for easy access 
when working; however, to obtain the maximum cooling duration, we recommend using the collar 
whenever possible. 

Quick Start
•  Remove two XT Cooling Cores from -20°C 

freezer  and  place  on  bench.                                                                

•  When temperature strip on XT Cooling 

Core registers 1°C, place CoolRack

®

 or 

CoolSink

®

 sample modules in the Cool-

Box

  XT  base  and  place  collar  on  top.                                       

•  Load samples in CoolRack or CoolSink 

modules.

Caution
•  Lift CoolBox 2XT from the bottom using 

the  handholds  on  the  sides.                                                                   

•  Always use two hands when carrying or 

lifting CoolBox 2XT.                                          

•  Avoid touching the top metal surface of 

the XT Cooling Core and XT Freezing Core 
when removing from freezer. 

IMPORTANT: To ensure your samples do not freeze and to get maximum cooling duration, please 
refer to the detailed instructions that follow.

*All tests were performed using a CoolRack

®

 XT M24 loaded with 24 TruCool

®

 2.0 mL microcentrifuge 

tubes filled with 1.5 mL water. Actual performance may vary depending upon CoolRack

®

 module em-

ployed, sample load, initial sample temperature, ambient temperature, air currents, and other conditions. 

Working 

Temperature Range

CoolBox XT 

Cooling Source

Temperature Duration

*

Open Lid

Temperature Duration

*

Closed Lid

0.5° to 4

o

C

XT Cooling Core

Over 10 hours

Over 16 hours

-20° to 0

o

C

XT Freezing Core

Over 5 hours

Over 8 hours

-78

o

C

200 ml Dry Ice

Over 4 hours

Over 5 hours

CoolRack

®

 and CoolSink

®

 Thermo-Conductive Sample Modules Compatible with CoolBox XT 

Item

Description

Accommodates

BCS-163

CoolRack M6

6 x 1.5 mL or 2.0 mL microfuge tubes

BCS-125

CoolRack M15

15 x 1.5 mL or 2.0 mL microfuge tubes

BCS-127

CoolRack M15-PF

15 x 1.5 mL conical microfuge tubes

BCS-535

CoolRack XT M24

24 x 1.5 mL or 2.0 mL microcentrifuge tubes, SBS-compliant

BCS-126

CoolRack CF15

15 x 1.0 mL or 2.0 mL cryogenic vials

BCS-534

CoolRack XT CFT24

24 cryogenic vials, with “gripping” wells for one-hand vial open/close, SBS-compliant

BCS-529

CoolRack XT PCR96

PCR tubes, strips or 96-well plate

BCS-523

CoolRack XT M-PCR

12 x 1.5 mL or 2.0 mL microcentrifuge tubes and 6 PCR strip wells

BCS-538

CoolRack XT PCR384

one 384-well PCR plate, SBS-compliant

BCS-231

CoolRack 96x0.5ml

96 x 0.5 mL 2D barcode tubes

BCS-149

CoolRack 96x1ml

96 x 1.0 mL 2D barcode tubes

BCS-536

CoolSink XT 96F

one flat-bottom 96- or 384-well assay plate, SBS-compliant

BCS-537

CoolSink XT 96U 

one u-bottom 96-well assay plate, SBS-compliant

Tall Tube Modules - requires Extension Collar, if lid closing is necessary

BCS-232

CoolRack L

12 x 15 mL centrifuge tubes

BCS-153

CoolRack 15ml

9 x 15 mL centrifuge tubes

BCS-154

CoolRack 50ml

4 x 50 mL centrifuge tubes

BCS-235

CoolRack LV

12 x 13mm or 16mm diameter tubes

BCS-157

CoolRack VS13

9 x 13x75mm blood tubes

BCS-155

CoolRack V13

9 x 13x100mm blood tubes or 5.0 mL cryogenic vials

BCS-156

CoolRack VS16

9 x 16x100mm blood tubes

CAUTION: The products described here are intended for the exclusive use by trained and experienced 

laboratory and medical personnel. Use of dry ice can be dangerous. Direct skin contact with dry ice or 
metal components that have been in contact with dry ice can cause freezing injury. Always use appropri-
ate protective equipment for eyes and skin when handling dry ice and cold metal components.

Made in USA.  