Troubleshooting Chart 

 

Problem 

Possible Causes 

Corrective Action 

Unit does not run 
(Digital Display is 
blank) 

No power to unit 
 

Refrigerated units only 

Electrical cord between 
Control Head and 
Refrigeration Power Module 
loose or unplugged   
Power Switch / Circuit 
Breaker in OFF position

 

Check that the electrical cord is secure and 
connected to an operating electrical outlet. 
 
Check that the electrical cord is secure and 
properly connected. 
 
 
Place Power Switch / Circuit Breaker in ON 
position. 

Unit does not run 
(STANDBY appears 
on Digital Display) 

Unit in Standby mode 

Press Power Key on front panel. 

No fluid circulation 

Insufficient fluid in reservoir 
Pump impeller jammed 

Add fluid to reservoir. 
Inspect pump and remove debris as required. 

Insufficient circulation 

Fluid viscosity too high 
External tubing diameter 
too small 
Low line voltage 

Replace with lower viscosity bath fluid. 
Replace with larger diameter tubing. 
 
Check and correct as required. 

Unit does not heat 

Insufficient fluid in reservoir 
Temperature set point too 
low 
Safety Set Temperature too 
low 

Add fluid to reservoir. 
Increase temperature set point. 
 
Increase Safety Set temperature. 

Insufficient heating 

Insufficient circulation 
Low line voltage 
Ambient temperature too 
cool 
Excessive heat loss 

See Insufficient circulation, above. 
Check and correct as required. 
Increase ambient temperature or relocate unit. 
 
Check for heat loss from external tanks and 
hoses; 
Check for vapor/heat loss from internal 
reservoir. 

Temperature 
unstable 

Insufficient circulation 
Debris or mineral build-up 
on pump, heater, or 
temperature sensor. 

Check pump flow and operation.  
Clean as required. 

Unit does not cool 

Dust build up on air filter or 
condenser 
Blocked air ventilation 
screens 
Temperature set point is 
too high 
Excessive heat load 
 
Ambient air temperature too 
high (>35°C / 95°F) 
Low or high line voltage 

Clean air filter and/or condenser as required. 
 
Remove blockages as required. 
 
Decrease temperature set point. 
 
Check that heat load does not exceed capacity 
of bath; correct as required. 
Decrease ambient air temperature. 
 
Check and correct as required. 

Содержание TC-150MX

Страница 1: ...TC Series Circulating Baths with MX Temperature Controller Operator s Manual 110 509 BEL EN 06 September 2018 Models TC 150MX TC 250MX TC 450MX TC 550MX TC 650MX...

Страница 2: ...ational Display 19 Set Up Sub Menus 20 Adjusting the Temperature Set Point 20 Selecting the Temperature Unit 21 Setting the Offset Calibration 22 Setting the Low Limit Temperature 23 Setting the High...

Страница 3: ...ump easy flow adjustment External circulation capability Fully enclosed housing prevents direct contact with pump and heater yet provides quick access for inspection and cleaning Suitable for use with...

Страница 4: ...lbs 29 9 kg Ambient 10 to 135 C Ambient 20 to 275 F TC 550MX Refrigerating Heating Circulator 7 liters 6 18 x 5 59 x 5 0 in 15 7 x 14 2 x 12 7 cm 23 2 x 16 2 x 17 3 in 58 9 x 41 1 x 43 9 cm 84 lbs 38...

Страница 5: ...TC 550 and TC 650 Heating Rates 60 Hz 50 Hz TC 550 and TC 650 Cooling Rates 60 Hz 50 Hz TC 150 TC 250 and TC 450 Heating Rates 60 50 Hz 0 20 40 60 80 100 120 140 0 20 40 60 80 100 120 Time Minutes Flu...

Страница 6: ...al dangers This symbol advises danger from electricity or electric shock This symbol indicates that a hot surface may be present This symbol marks information that is particularly important This symbo...

Страница 7: ...efore performing any maintenance or service WARNING Never operate the Circulator without bath fluid in the reservoir Periodically check the reservoir to ensure that the liquid depth is within acceptab...

Страница 8: ...Directive 2006 95 EC EC Electromagnetic Compatibility Directive 2004 108 EC IEC 61010 1 2001 IEC 61010 2 2001 IEC 61326 2005 EN 61326 2006 Unpacking Your Circulator Your Circulator is packed in a spec...

Страница 9: ...1 2 ID Black Silicone Hose 3 Long 2 pcs 1 2 HB x 1 4 HB GF Nylon Fitting 2 pcs Bypass Tubing with Spring PS510 713 Latex Tubing 6 ft 0 25 ID EX Tubing Maximum temperature 80 C Beaker Platform s for B...

Страница 10: ...nd Components MX Controller Backlit Digital Display Up Arrow Button Down Arrow Button Menu Button Power Button Safety Thermostat under cover plate Actual Bath Temperature Set Point Temperature Heating...

Страница 11: ...lter behind access panel Drain Valve and Port right side on model TC 550MX Thermometer Port Power Switch Circuit Breaker IEC Power Connection to Mains Refrigeration Control Cable Connection IEC Power...

Страница 12: ...deck Always disconnect electrical power and drain fluid from bath before moving WARNING To avoid the potential for burns allow the Circulator to cool completely before cleaning or performing any maint...

Страница 13: ...external water line Use of cooling coils is optional 3 Connect all power cords and control cables Maximum 1 in 2 54 cm below the underside of the top deck Minimum 2 6 in 6 7 cm below the underside of...

Страница 14: ...Power Switch Circuit Breaker ON Refrigerating Heating Baths only 5 Turn Controller ON 6 Enter temperature set point 7 Set safety thermostat once unit reaches set point See Installation and Startup Con...

Страница 15: ...carefully before filling reservoir WARNING See Technical Information in the rear of this manual for a list of compatible liquids WARNING If the proper fluid level is not maintained the heater coil may...

Страница 16: ...rear of the MX Controller Use the fittings provided with the unit to make external connections If a higher flow rate is desired you can move the flow adjuster toward a closed position as shown below A...

Страница 17: ...to potential reduction in flow rate of circulating fluid which may ultimately affect temperature stability at the external device Refrigeration Control Connections Refrigerating Heating Circulators o...

Страница 18: ...emale IEC connector on the Refrigeration Power Module Attach the 6 ft 1 8 m power cord to the IEC electrical connection on the Refrigeration Power Module and then plug the male connector into the Main...

Страница 19: ...e until it stops 2 Press the or key SET will begin flashing Continue pressing the and keys until the set point temperature is equal to your desired Safety Set temperature SET will stop flashing approx...

Страница 20: ...Turning Your Circulator ON Press the key All characters symbols on the LCD will momentarily light When the Circulator begins running the actual and set point temperatures will be displayed and the cir...

Страница 21: ...Press or To accept a value in a sub menu press or allow the display to time out approximately 10 seconds Adjusting the Temperature Set Point This is the temperature at which the fluid in your Circulat...

Страница 22: ...e actual bath temperature and set point temperature are displayed The factory default is C To Access Press the key until C F appears on the display To Change Press to select C press to select F To Acc...

Страница 23: ...below Heat Only Circulators 1 Remove the power cord from the electrical outlet 2 Plug the power cord back into the electrical outlet while pressing and holding the key 3 When STANDBY appears on the d...

Страница 24: ...Limit value may be set from 35 to 10 C 30 to 50 F the factory default is 35 C 30 F To avoid unwanted alarms or shutdowns during regular operation the Low Limit value should be set at least 5 lower tha...

Страница 25: ...mit value may be set from 40 to 140 C 100 to 285 F the factory default is 140 C 285 F To avoid unwanted alarms or shutdowns during regular operation the High Limit value should be set at least 5 highe...

Страница 26: ...ature is recommended The Auto Cool range is from 1 to 135 C The factory default is 45 C Conventional Refrigeration TC 550MX and TC 650MX Refrigerating Heating Circulators use a conventional refrigerat...

Страница 27: ...e Operational Parameter Factory Default Value Temperature Scale C Temperature Set Point 20 C 68 F Calibration Offset 0 0 C Low Limit 35 C 30 F High Limit 140 C 285 F Auto Cool 45 C Loss of Power Resta...

Страница 28: ...mpressor and pump will remain OFF until the problem is corrected Allow bath to warm or add heat load Decrease the Low Limit temperature value Alarm The bath temperature has risen above the High Limit...

Страница 29: ...nt algae contamination and minimize the frequency of draining the reservoir an algaecide such as TC Fluid 1A should be used WARNING Do not use chlorine bleach Draining the Bath Reservoir WARNING Bath...

Страница 30: ...pend on the size of the bath and the difference between the initial bath temperature and the Safety Set temperature 2 Once the bath temperature has stabilized slowly rotate the Safety Thermostat count...

Страница 31: ...mperature Controller s wetted parts The cleaner should be added to the bath reservoir at the prescribed dosage and circulated at 60 C 140 F until the scale is removed External Surfaces Only mild deter...

Страница 32: ...from the top deck without the use of special tools It is removed as follows 1 Place the tip of a small flat blade screwdriver under the retaining ring locking tab and pry up gently 2 Rotate the Temper...

Страница 33: ...one another allowing you to position the Temperature Controller at either of two different viewing angles 1 With the retaining ring locking tab oriented above one of the indents on the top deck slowl...

Страница 34: ...required Unit does not heat Insufficient fluid in reservoir Temperature set point too low Safety Set Temperature too low Add fluid to reservoir Increase temperature set point Increase Safety Set tempe...

Страница 35: ...and correct as required Unable to achieve low end extreme temperatures Pump speed too high Incorrect bath fluid Insufficient insulation on external fluid lines Ambient air temperature too high 35 C 9...

Страница 36: ...90 C distilled water is recommended For temperatures below 10 C a mixture of laboratory grade ethylene glycol and water should be used Do not use deionized water The following chart is intended to ser...

Страница 37: ...reme The presence of ice or slush adversely affects temperature stability A viscosity above 10 centistokes adversely affects temperature uniformity A high fluid viscosity and high pump speed adds heat...

Страница 38: ...ity for separate collection and recycling It is also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect the persons in...

Страница 39: ...ir Filter TC 650 PS305 054 O Ring for Drain Valve TC 550 and TC 650 baths PS400 934 Accessory connection Hose rated for 50 to 80 C EX Tubing Refrigeration Control Cable TC 550 and TC 650 PS525 879 Res...

Страница 40: ...S A Telephone 508 946 6200 FAX 508 923 5009 www brookfieldengineering com For repair or service outside the United States consult Brookfield Engineering Laboratories Inc or the dealer from whom you p...

Страница 41: ...110 509 BEL EN 40 Manufactured by PolyScience 6600 W Touhy Avenue Niles IL 60714 U S A 1 800 229 7569 1 847 647 0611 www polyscience com...

Отзывы: