Inaccurate Readings
❏
Verify spindle, speed and model selection
❏
Verify test parameters: temperature, container, volume, method. Refer to:
• “More Solutions to Sticky Problems”; Section II.2A - Considerations for Making Mea-
surements
• Dial Viscometer Operating Manual; Appendix B - Viscosity Ranges
• Dial Viscometer Operating Manual; Appendix C - Variables in Viscosity Measurement
❏
Perform a calibration check. Follow the instructions in Appendix D.
• Verify tolerances are calculated correctly.
• Verify calibration check procedures were followed exactly.
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix G for
details on how to return your viscometer.