Brookfield Engineering Labs., Inc. 

Page 3 

Manual No. M02-33E209

Spindle 04
CALIBRATE?  

YES/NO

Figure II-4

If you are using the CAP Viscometer for the first time or have just received the instrument back 

from being serviced, press YES and see Section III.  Cones supplied at the time of order are 

calibrated to the viscometer by Brookfield prior to shipment, but should still be verified with a 

calibration check prior to use for the first time.  Cones purchased separately at a later time by 

the user must also be calibrated prior to making viscosity measurements.  If you have 2 cone 

spindles that are identical (have the same number), you must do a cone calibration when you 

swap them.

Otherwise, there is no requirement to perform a cone calibration.  

Note

:

    1.  CAP Viscometers allow for 

only one

 cone at a time of the 

same cone number

 to 

be calibrated to the viscometer.  Multiple cones of the same cone number 

must each

 

be calibrated

 to the viscometer before operation (refer to 

Section III

).

 

 

2.  A special feature of the CAP series viscometers allows the user to perform a cone 

calibration with viscosity standard fluids.

 

  3.  Cones entered as 11 through 99 must be first calibrated following the directions in 

Section III.

If you are not going to calibrate the cone, continue by pressing the 

NO 

key, then the 

ENTER 

key.  

The viscometer will display the 

MAIN

 

SCREEN

 (

Figure

 

II-1

).  Should you subsequently run 

a test and observe that the instrument is reading all “0s”, the “P” for Poise is flashing, and the 

“%” symbol is flashing, then the cone spindle you are using is not calibrated.  You must perform 

the cone spindle calibration to rectify the problem.

If you are going to calibrate the cone, press the 

YES 

key, refer to 

Section III

 and follow the 

calibration instructions under 

Cone Calibration

.

 II.3 Speed Setting

The CAP 2000+ is a variable speed viscometer.  The speed of rotation is shown in the lower 

right corner of the display.  To change the speed of rotation, enter the new speed using the 

number keys.  The new speed will be shown in the lower right hand corner of the display.  To 

accept the new speed, press 

ENTER

.

The new speed may be cancelled by pressing 

ESCAPE 

before pressing 

ENTER

.

To begin spindle rotation, press the 

RUN 

key.

Note:

  The speed cannot be changed while the motor is running.

Speed control may be configured to either shear rate or rotations per minute (see Special Func

-

tions in section II.1).  Data entry for speed selection is the same in either configuration.

Data entry must be in whole numbers.

 

Содержание CAP 2000+H

Страница 1: ...946 6200 FAX 508 946 6262 or 800 628 8139 USA excluding MA INTERNET http www brookfieldengineering com BROOKFIELD ENGINEERING LABORATORIES INC 11 Commerce Boulevard Middleboro MA 02346 USA with office...

Страница 2: ...top Keys 15 II 9 Parameter Display 16 III OPERATION 17 III 1 Full Scale Range and Accuracy of Measurement 17 III 2 Accuracy of Viscosity and Temperature 18 III 3 Calibration Verification 20 III 4 Cone...

Страница 3: ...er CGS or SI units CGS SI Comment Viscosity P or cP Pa s or mPa s 0 1 Pa s 1 P 100 cP Shear Rate Sec 1 Sec 1 Speed RPM RPM Temperature C C The CAP 2000 Viscometer outputs data to a parallel printer in...

Страница 4: ...p P N C1K 63 Column Head Serial Tag Info on Back of Viscometer Head Base Console Cone Spindle P N CAP S XX The Flat Thumb Screw P N C1K 34Y Figure I 1 Components I 2 Utilities Input Voltage 115 VAC or...

Страница 5: ...g Net Weight 27 lb 12 3 kg Carton Volume 4 9 cu ft 0 15 m3 Carton Dimensions 18 in L x 18 in W x 26 in H 48 cm L x 48 cm W x 66 cm H Materials CAP cone spindles and temperature plates are made of tung...

Страница 6: ...Brookfield Engineering Labs Inc Page Manual No M02 313E1209 I 4 Dimensional Details...

Страница 7: ...ing CAP Viscometer Figure I 2 Detail of Foam Insert 3 Verify that the viscometer s power requirements match your power source BEFORE con necting it to power TheAC input voltage and frequency must be w...

Страница 8: ...ace Refer to the manual for specific warning or caution information to avoid personal injury or damage to the instrument Safety Overview If this instrument is used in a manner not specified by the man...

Страница 9: ...This key selects time entry mode wait time before spindle rotates SPINDLE This key selects the cone spindle entry mode TEMP This key selects the temperature entry mode I 8 Viscosity and Temperature D...

Страница 10: ...brasive cloth and solvent appropriate for sample material that is not aggressive to immersed compo nents Do not use metal objects to clean the plate surface as scratching of the plate may oc cur and c...

Страница 11: ...f measure and speed control may be selected through the special functions screen This screen is accessed by pressing the STOP key during instrument power up The CAP 2000 can be configured to control s...

Страница 12: ...he slot by the thumb screw and turning the trap clockwise onto the thumbscrew Slide the cone up into the adapter as far as it will go and hand lock it in place with the thumb screw Tighten the thumb s...

Страница 13: ...must be first calibrated following the directions in Section III If you are not going to calibrate the cone continue by pressing the NO key then the ENTER key The viscometer will display the MAIN SCRE...

Страница 14: ...After inserting the sample material onto the plate lower the spindle and solvent trap and allow sufficient time for thermal equilibrium prior to starting the test II 5 Hold Time Settings Hold time se...

Страница 15: ...D CM2 S RATE 1 sec SPEED RPM TIMER SEC CONE No 25 0 10000 0750 20 02 Figure II 5 A maximum of 999 seconds can be printed when running in manual TIMER mode 00 Over 999 seconds will print EEE The TIMER...

Страница 16: ...ed All relevant measurement parameters will be shown for 5 seconds including speed shear rate run time hold time and spindle The display will return to the default screen after five seconds Note The v...

Страница 17: ...torque model you have consult the se rial tag of your viscometer or the certificate of calibration that accompanied the instrument or the display on power up 23CAP 2000 Low Torque CAP 2000 High Torqu...

Страница 18: ...ts Cone range constant 22 7 N FSR Poise where N RPM For High Torque CAP instruments Cone range constant N FSR Poise where N RPM The last column in the above tables shows this calculation III 2 Accurac...

Страница 19: ...100 FSR 50 FSR 50 FSR 01 2 0 2 0 2 0 4 0 02 2 0 2 0 2 0 4 0 03 2 0 2 0 2 0 4 0 04 2 0 3 0 3 0 6 0 05 2 0 4 0 4 0 8 0 06 2 0 5 0 5 0 10 0 07 2 0 08 2 0 09 2 0 10 2 0 2 0 2 0 2 0 Not recommended for us...

Страница 20: ...h cone spindle to the CAP Viscometer See Section III 4 Calibration verification of your viscometer should also be performed when viscosity readings with your product are suspect Verifying the calibrat...

Страница 21: ...your standard If your measured viscosity falls within this window the verification passes Example CAP 03 spindle on High Torque Low Temp CAP Viscometer running at 900 RPM Viscosity standard has nomin...

Страница 22: ...ndles on High Torque CAP 2000 HIGH TORQUE LOW TEMP HIGH TORQUE HIGH TEMP Cone Fluid Part Num ber Nomi nal Value cP Temp C Sample Size micro liters Cone Fluid Part Num ber Nomi nal Value cP Temp C Sam...

Страница 23: ...or CALIBRATE and press ENTER 9 Enter the appropriate values for temperature C and viscosity cP value printed on the label Note Viscosity values are always entered in units of CENTIPOISE cP no matter w...

Страница 24: ...___________________ Fluid _____________________________________________________ Figure III 3 III 5 Repeatability The CAP 2000 Viscometer is repeatable to 0 5 of the full scale viscosity range FSR Due...

Страница 25: ...the containment of solvents and or prevention of sample drying The trap should be used for all measurements with high tempera ture instruments CAP 2000 H d If a printer is to be used it should be con...

Страница 26: ...AP H series Viscometers the tem perature display may indicate a deflection from the equilibrium temperature setting as the cone begins rotating at high shear rates The temperature display may indicate...

Страница 27: ...f the sample plate for integrated viscosity temperature tests between 5 C and 75 C CAP 2000 L or 50 C and 200 C CAP 2000 H depending on viscometer model CAPCALC features include on screen and printed...

Страница 28: ...e fluids have different viscosities at different shear rates different RPMs They fall into two groups 1 Time Independent non Newtonian 2 Time Dependent non Newtonian The time dependency is the time th...

Страница 29: ...pm and a second at 900 rpm the two viscosity values produced may be different because the readings were made at different shear rates The faster the spindle speed the higher the shear rate The shear r...

Страница 30: ...he serial cable The cable connections are Com Port RxD pin 2 9 pin or pin 3 25 pin to CAP Txd pin 3 Com Port TxD pin 3 9 pin or pin 2 25 pin to CAP RxD pin 2 Com Port ground pin 7 to CAP Serial Ground...

Страница 31: ...eading zeros cc cone All cc values shall be padded to 2 characters with leading zeros I CR ICAP bbxxxddddd ss CR Identify the viscometer and firmware in use bb HI or LO temperature range xxx firmware...

Страница 32: ...re multiplied by 100 before transmission from the CAP 2000 All FSR values are four characters padded with leading zeroes rrrrrr Shear rate 1 seconds values hexadecimal Shear rate values are multiplied...

Страница 33: ...indicating a high temperature CAP 2000 with Version 2 20 firmware spring constant of 1 00 and a status byte of 0 3 The host sends a T190 The CAP 2000 begins controlling to 40 0 C and responds with T1...

Страница 34: ...46 U S A Telephone 508 946 6200 FAX 508 923 5009 www brookfieldengineering com For repair or service outside the United States consult Brookfield Engineering Laboratories Inc or the dealer from whom y...

Страница 35: ...e available for immediate shipment from Brookfield If the case is not available take care to wrap the instrument with enough material to support it Avoid using foam peanuts or shredded paper R Fill ou...

Страница 36: ...rol _________________ Sample Temperature _______________________________ Accessories __________________________________________________________ SERVICE AND RETURN INSTRUCTIONS Please check service s r...

Отзывы: