Bronkhorst®
Instruction Manual mini CORI-FLOW ML120
9.17.097G
33
Customer Model
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[16]
RW
-
93
113/4
0xF120…0xF127/ 61729…61736
This parameter is used to add extra information to the model number information, such as a customer-specific model
number.
Serial Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[20]
R
-
92
113/3
0xF118…0xF11F/ 61721…61728
Instrument serial number for identification.
BHT Model Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[27]
RW
-
91
113/2
0xF110…0xF117/ 61713…61720
This parameter shows the Bronkhorst® instrument model type information.
Firmware Version
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
105
113/5
0xF128…0xF12A/ 61737…61739
Revision number of the firmware
Identification Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
175
113/12
0x0E2C/3629
Bronkhorst® (digital) device type identification number.
Device Type
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
90
113/1
0xF108…0xF10A/ 61705…61707
Device type information string; this parameter contains an abbreviation referring to the identification number.
4.3
Measurement and control
Basic
Measure
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
R
0…41942
(65535*)
8
1/0
0x0020/33
This parameter indicates the flow metered by the instrument. The signal of 0…100% is presented in a range of 0…32000.
The maximum measured value output is 131.07%, which translates to 41942.
*In case the instrument is prepared for bi-directional measurement, the negative signals with an output range of -
73.73…-0.003% are represented by the range of 41943…65535, whereas the positive signals 0…131.07% are still
represented by the range of 0…41942. (FlowDDE converts the numbers to negative values automatically).