Bronkhorst®
Instruction Manual MI Series Industrial mini CORI-FLOW™
9.17.120A
34
Available modes:
Value
Description
0
Counter off (default)
1
Counting up continuously
2
Counting up until limit reached (set by
Counter Limit
)
Counter Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[4]
RW
see table
below
128
104/7
0xE838…0xE839/59449…59450
This parameter contains the name of the counter readout unit.
Counter Unit
supports the following values:
Mass
Normal volume
(1.01325 bar(a), 0 °C)
Standard volume
(1.01325 bar(a), 20 °C)
Custom volume
(
Capacity Unit Pressure
,
Capacity Unit Type
Temperature
)
ug, mg, g, kg
uln, mln, ln,
mm3n, cm3n, dm3n, m3n
uls, mls, ls,
mm3s, cm3s, dm3s, m3s
ul, ml, l,
mm3, cm3, dm3, m3
Counter Value
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…
10000000
122
104/1
0xE808…0xE809/59401…59402
Current counter value in units selected with parameter
Counter Unit
.
Counter Limit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW
0…9999999
124
104/3
0xE818…0xE819/59417…59418
Counter limit/batch size in units selected with parameter
Counter Unit
.
Default value: 0.
Counter Setpoint Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…1
126
104/5
0x0D05/3334
Specifies whether or not to change the setpoint after reaching the counter limit.
Value
Description
0
No setpoint change (default)
1
Change setpoint to
Counter new setpoint
Counter New Setpoint
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned int
RW
0…32000
127
104/6
0x0D06/3335
New (safe) setpoint when a counter limit is reached until reset. Range 0…32000 represents 0…100% setpoint.
Default value: 0