Bronkhorst®
Instruction Manual FLEXI-FLOW™ Compact
9.17.158A
34
Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
R
0…7
114
115/8
0x0E68/3689
This parameter is used to reset the program, counter or alarms (alarms are reset on all channels).
Value
Description
0
No reset
1
Reset counter
2
Reset alarm
3
Reset counter
4
Reset and disable counter
5
Reset firmware program (soft reset)
6
Reset
Alarm info
error bit
7
Reset
Alarm info
warning bit
·
Resetting alarms or the counter can be disabled by
Reset Alarm Enable
or
Reset Counter Enable
respectively.
·
Reset
value 2 resets alarms on all three communication channels, unless resetting is disabled for 1 or more channels.
Wink
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char [27]
W
0…9*
1
0/0
0x0000/1
Sending any text string value between 1 and 9 to this parameter makes the status indication light flash white for that
number of seconds. This can be useful in order to identify a specific device in a large fieldbus network.
*) Modbus requires the following values:
No. of seconds
Wink value
1
12544
2
12800
3
13056
4
13312
5
13568
6
13824
7
14080
8
14336
9
14592
Control Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
12
115/1
0x0024/37
Control Mode
is used to select different control modes of the instrument and determines from which source(s) it accepts a
setpoint. The following control modes are available:
Value
Mode
Instrument action
Setpoint source
0
BUS/RS232
Controlling
Fieldbus/RS-232
2
FLOW-BUS Slave
Acting as slave instrument on FLOW-BUS
FLOW-BUS master output x
Slave Factor/
100%
3
Valve Close
Controller disabled, valve closed
4
Controller Idle
Controller disabled, valve frozen in current position
7
Setpoint 100%
Controlling, setpoint fixed to 100%
8
Valve Fully Open
Controller disabled, valve fully opened
9
Calibration Mode Calibration mode enabled
(factory only)
12
Setpoint 0%
Controlling, setpoint fixed to 0%
18
RS232
Controlling, safe state deactivated
Fieldbus/RS-232