Bronkhorst®
Instruction Manual EL-FLOW® Select
9.17.099J
36
Value
Mode
Instrument action
Setpoint source
7
Setpoint 100%
Controlling, setpoint fixed to 100%
8
Valve Fully Open
Controller disabled, valve fully opened
9
Calibration Mode Calibration mode enabled
(factory only)
10
Analog Slave
Acting as slave of other instrument in analog mode
Analog Input
x
Slave
Factor/
100%
12
Setpoint 0%
Controlling, setpoint fixed to 0%
13
FLOW-BUS
Analog Slave
Acting as slave of other instrument on FLOW-BUS, slave
factor is set by analog input signal
RS-485 only: FLOW-BUS
master output x
Analog Input
18
RS232
Controlling, safe state deactivated
Fieldbus/RS-232
20
Valve Steering
Controller disabled, setpoint redirected to
Valve output
21
Analog Valve
Steering
Controller disabled, analog input redirected to
Valve output
22
Valve Safe State
Force instrument in
·
The default value is either 'Analog input’ or ‘BUS/RS232’ (depending on the ordered default setting).
·
If
Control mode
is changed to value 0, 1, 9 or 18, the instrument returns to the default control mode at the next power-up
or reset. Other values are retained after power-up or reset.
Calibration Mode
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
B
0, 9, 255
58
1/4
0x0E61/3682
After enabling calibration mode by means of parameter
Control Mode
, this parameter is used to start the autozero function
of the flow sensor. The following modes are supported:
Value
Description
0
Idle (no action)
9
Start zeroing
255
Error (result of previous calibration mode)
4.1.1
Default control mode
IO Status
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
B
0…255
86
114/11
0x0E4B/3660
The instrument is set to accept a setpoint from either an analog or a digital source. Although this setting can be changed
with parameter
, the instrument usually returns to its default control mode at every power-up or reset. The
default control mode can be set with parameter
IO Status
; to change it, use the procedures as described below.
Changing from digital operation to analog operation:
1. Set parameter
Init Reset
to 64 (unlocked)
2. Read parameter
IO Status
3. Add 64 to the read value
4. Write the new value to parameter
IO Status
5. Set parameter
Init Reset
to 82 (locked)
Changing from analog operation to digital operation:
1. Set parameter
Init Reset
to 64 (unlocked)
2. Read parameter
IO Status
3. Subtract 64 from the read value
4. Write the new value to parameter
IO Status
5. Set parameter
Init Reset
to 82 (locked)
The procedures described above do not change the value of parameter
Control Mode
directly. To apply the new default
control mode immediately, change the value of parameter
Control Mode
manually or reset or restart the instrument.