Extended Defrosting
The evaporator (cooling) coil of the refrigerator may build up a layer of ice over time. This will
reduce the performance of the unit – inspect the evaporator coil for ice build-up regularly.
For extended defrosting, turn o
ff
the unit with the door opened for a period of 24 hours.
Condenser Coil
IMPORTANT:
The condenser coil must be cleaned with a vacuum or brush once a
month. The condenser coil must be cleaned regularly to maintain the performance
and minimise power consumption. A poorly maintained condenser coil can lead to
failure of the unit and cabinet temperatures above 8°C.
7. MAINTENANCE
Cleaning the Cabinet
• Remove the products and transfer to a suitable location prior to cleaning the inside of the
cabinet.
• Do not use any aggressive or abrasive cleaning agents.
• Never use a water hose or a high-pressure jet to clean.
• Do not use sharp or pointed objects for cleaning.
• Clean the inside and outside of the appliance with a clean damp cloth or with a neutral
cleanser.
• Dry all surfaces with a dry cloth.
WARNING
: Disconnect the appliance from the power supply before performing any
maintenance. Never unplug the appliance by pulling on the cord. Always grasp the
plug firmly and pull straight out of the outlet.
WARNING
: Exposure to significant water ingress will void warranty and cause
potential safety issues. If the unit should be exposed to water, please contact the
manufacturer immediately.
IMPORTANT: Warranty may be void if there is no evidence of preventative
maintenance is performed on the appliance.
1300 276 642
www.bromicrefrigeration.com.au
10
6. PRODUCT LOADING AND STORAGE
Start-up
• Clean the interior surface with a soft cloth and warm water before loading.
• Ensure there is free circulation of air between the products.
• Do not block the evaporator fan.
• Do not overstock the refrigerator.
• Use cold mass (i.e. water bottle) on unused shelves.
• Keep refrigerator door openings to a minimum.
• Ensure products are between 2 and 8°C before loading – if there is a breach in the cold-
chain, isolate the a
ff
ected products. Do not use or discard until advise is given.
• Some products are sensitive to ultraviolet (UV) light, please see storage guidelines from
the manufacturer before removing from packaging.
• Do not store food and other goods in the refrigerator.
• Maximum recommended shelf load is 20kg.
NOTE: Bromic cannot be responsible for loss of stock or damage to stock.
Please refer to the ‘National Vaccine Storage Guidelines: Strive for 5’ for more information on
vaccine storage and handling or contact your local state or territory health department.
(http://www.health.gov.au/internet/immunise/publishing.nsf/content/IMM77-cnt)
WARNING
: Failure to follow the guidelines below will result in cabinet
temperatures above 8°C
IMPORTANT
: Run the refrigerator for at least 2 hours before loading.