Although measurement transducers are not specifi -
cally included in the MID regulations as they do not
form a complete measuring (system) instrument ref
Article 1 and 4, Annex I and Annex MI-005.
Brodie Meter Co., LLC implements the same stringent
regulations for all products and tests to the same
standards which are used for complete (systems)
instruments.
The complete system must contain all the necessary
components to meet the requirements of the local
regulations. These components may include, pumps,
air eliminators, strainers, valves, fl ow computers, etc.
The unit must be sealed in accordance with the local
regulations; it is the end users responsibility to ensure
this happens.
Flow measuring devices are provided with two labels
which specify fl ow ranges. The name plate label which
includes the factory serial number; details the op-
erating fl ow range, this is the fl ow range the device
will operate within without causing damage, and the
custody transfer label; this label details the working
fl ow range associated with a particular weights and
measures approval.
It should be noted that these may not be the same;
therefore in trade applications the fl ow ranges speci-
fi ed on the custody transfer label should be followed.
Essential Instructions for Measuring Equipment Including the Euro-
pean Union (Directive 2004/22/EC MID)
Essential Instructions for Electrical Equipment Including the Euro-
pean Union (Directive 2004/108/EC and 2004/22/EC)
This unit contains Electrostatic sensitive circuit
boards. Electrostatic safety precautions should be
taken to prevent damage.
When connecting wiring it is good practice to use
shielded cable. The shield should be connected to
earth at the read out or control systems end of the
cable; the other end of the shield should not be con-
nected.
This wiring practice is mandatory in order to comply
with the requirements for electromagnetic compati-
bility as per the EMC directive 2004/108/EC and MID
2004/22/EC of the council of the European Union.
It is the end users responsibility to ensure that all pro-
tective covers are in place to prevent electrical shock
and/or personnel injury.