10 Servicing and Calibration
IMPORTANT: THE HEATER AND CONTROL SYSTEM ARE AT MAINS VOLTAGE. NEVER ATTEMPT ANY KIND OR
FORM OF SERVICING UNLESS THE MACHINE IS DISCONNECTED FROM THE MAINS ELECTRICITY SUPPLY.
RISK OF ELECTRIC SHOCK!
10.1 Under certain conditions it is possible that condensation may form on the inner walls. The presence of
water gathering at the base of the Hatcher does not affect the performance of your Hatcher and does
not pose an electrical hazard.
10.2 In case of a failure first check that the mains power supply is working and that the mains cable
connector is fully engaged in the socket on the rear panel. The digital control system may be reset to
the original factory defaults by connecting the power supply while holding the OK button. Check
temperature calibration after resetting to defaults.
10.3 If the problem persists contact your distributor or Brinsea Products Service Dept.
10.4 The functional parts of the Ova-Easy Advance Series II incubators are modular and parts are available
and are readily exchanged by a suitably qualified person equipped with basic tools. Fitting instructions
are supplied with replacement parts.
10.5 The digital temperature and humidity display is individually calibrated during manufacture but may be
re-calibrated if required. It is NOT RECOMMENDED that this procedure is carried out by the user.
BE CAUTIOUS OF LOW COST ANALOGUE OR DIGITAL THERMOMETERS AND HYGROMETERS.
BRINSEA PRODUCTS LTD USES SOPHISTICATED EQUIPMENT TRACEABLE TO INTERNATIONAL
REFERENCE STANDARDS.
To access the Calibration Menu press all three buttons simultaneously to unlock the display.
PRESS ALL 3 BUTTONS TO UNLOCK THE CALIBRATION MENU
SELECT THE OPTION / RETURN TO THE MENU.
GO FORWARD ONE SCREEN / INCREASE THE VALUE.
GO BACK ONE SCREEN / DECREASE THE VALUE.