6.3

 

Change paper towel at each feed or at least four times a day. 

 
6.4 

Follow feeding regimes recommended for your species ensuring the highest standards of 
hygiene at all times. 

 
6.5 

In order to maintain maximum air-flow the air-flow filters should be checked once a week for 
dust or down, and cleaned if necessary. 

 
6.6  

The TLC-4 can be used as a thermal intensive care unit for mammals or reptiles. In such cases 
fit the unit with the floor liner accessory (available from Brinsea Products). The unit would 
normally be set at between 30 and 35°C for chilled or shocked small animals. Brinsea do a 
medical version of the TLC-4 (the TLC-4M and the larger TLC-5M) which have further 
medical features for veterinary applications.  Contact Brinsea on 0845 2260120 or 

www.brinsea.co.uk

 for details. 

 
 
7.0

 

Cleaning Up

  

 

 

 

IMPORTANT: 

 

ALWAYS DISCONNECT THE TLC-4 BROODER FROM THE MAINS POWER SUPPLY 
DURING CLEANING. 

 
 

 ENSURE THAT ALL ELECTRICAL PARTS ARE KEPT DRY. 

 

7.1 

Following each brood in your TLC-4 brooder remove all debris from the floor. Wipe all 
internal surfaces with a soft cloth soaked in 100:1 Brinsea Incubation Disinfectant solution. 
Filters should be inspected on a weekly basis and be cleaned in necessary.  Remove both 
filters and gently hand wash in warm water then allow to dry before use. Filters need replacing 
every six months. Immerse and soak the nest tubs in disinfectant solution.   The exterior of the 
TLC-4 may be cleaned with a damp cloth.  

 

USE WATER BASED DISINFECTENTS ONLY WHEN CLEANING 

 

7.2 

 Always clean the TLC-4 brooder before storage and ensure that the unit is totally dry inside 
and out. 

 

7.3 

Re-order Brinsea Incubation Disinfectant, filters or evaporating block by phoning 0845 
2260120 

 
  
8.0  

Servicing

 

 

IMPORTANT. 

THE HEATER IS AT MAINS VOLTAGE. NEVER DRILL INTO OR PUNCTURE 

THE CURVED SIDES OF THE INCUBATOR.   RISK OF ELECTRIC SHOCK. 

 

8.1 

 Replace both filters after every six months of use. Contact Brinsea Products on  0845 2260120 
to order. 

 

8.2 

 The interior light bulb can be replaced if necessary.  Spares are available from Brinsea 
Products.  
 
ENSURE THAT THE TLC-4 IS DISCONNECTED FROM THE MAINS SUPPLY.  
 
Remove the twelve screws that retain the ceiling and replace the bulb with 15Watt pygmy BC 
type. Do not exceed the recommended Wattage. 
 

Содержание Octagon TLC-4

Страница 1: ...s For Brooding Birds and other Altricial Species Contents Section Subject Page 1 Introduction 3 2 Unpacking 4 3 Location and Installation 4 4 Temperature 5 5 Humidity 6 6 Introducing your Animals and Birds 6 7 Cleaning up 7 8 Servicing 7 9 Upgrading 7 10 Specifications 8 ...

Страница 2: ...sed and the user is advised to arrange his own insurance cover where loss of power or mechanical or electrical failure might result in unacceptable losses It is not recommended that animals of significant value be housed in this product unless it is used in conjunction with an independent temperature alarm system Such systems are available from Brinsea Products Ltd ...

Страница 3: ... books and veterinarian texts available covering animal and avian hand rearing and recovery techniques The principle applications for the TLC 4 are as an advanced brooder for altricial hatchlings Care must be taken to ensure adult birds do not cause damage to the brooder interior particularly interference with the filter or components above the filter Damage to these parts could cause injury or de...

Страница 4: ...nt if several units are in use at the same time Ensure that the room temperature cannot drop on a cold night Ideally thermostatically control the room at between 20 and 27 C 68 and 80 F Never allow the room temperature to drop below 15 C 59 F and ensure that the TLC 4 cannot be exposed to direct sunlight 3 1 Place the TLC 4 on a flat level surface workbench height is ideal with the door facing for...

Страница 5: ...lise throughout the unit before making adjustments to achieve the desired temperature 4 6 To adjust temperatures rotate the knob on the right hand side of the control panel clockwise to increase temperature anticlockwise to reduce it Follow the approximate guide below to achieve desired temperature settings Always allow half an hour between adjustments for the temperature to stabilise and refer to...

Страница 6: ...midity level RH and can cause dehydration A water reservoir is fitted to counteract this effect 5 1 Your TLC 4 brooder is fitted with a water reservoir see fig 1 which humidifies air as it is drawn into the brooder through the air inlet filter Use a solution of Brinsea Incubation Disinfectant 1part concentrate to 100 parts water in the water reservoir to inhibit bacterial build up It is recommende...

Страница 7: ...m the floor Wipe all internal surfaces with a soft cloth soaked in 100 1 Brinsea Incubation Disinfectant solution Filters should be inspected on a weekly basis and be cleaned in necessary Remove both filters and gently hand wash in warm water then allow to dry before use Filters need replacing every six months Immerse and soak the nest tubs in disinfectant solution The exterior of the TLC 4 may be...

Страница 8: ... includes medical extras such as a Nebuliser fitting oxygen supply fitting and temperature alarm input socket For further details on the product and upgrade call 0845 2260120 or visit www brinsea co uk 10 0 Specification Construction Top and base mouldings structural insulated polyurethane Curved sides Twin skin ABS Door opening frames Anodised extruded aluminium Clear door Plexiglass or similar s...