background image

 

20 

(4) Command: T<CR>    (54h 0dh) 

Response: TARE function is activated, and then returns scale status. just like 
pressing 

TARE

 button: 

<LF> H1H2H3<CR><ETX> 

      If TARE function cannot be activated, it will return to current scale status. 
(5) Command: U<CR>    (55h 0dh) 

Response: Changes units of measure and return scale status with new units, 
just like pressing 

UNIT

 button. The new measure unit should be allowed to 

use as per P11 setting.   
<LF>u1u2<CR><LF> H1H2H3<CR><ETX> 
If the weight unit is lb:oz,    U1U2= lb oz

 

(6) Command: L<CR>    (4ch 0dh) 

Response: If Hold function can be activated, it will enable/disable hold (lock) 
function, like the 

HOLD

 key is pressed, and returns scale status. 

<LF> H1H2H3 <CR><ETX> 

(7) Command: X<CR>    (58h 0dh) 

Response: power off the scale, just like press down the 

ON/OFF

 key to turn 

off the scale. 

(8) Command: all others 

Response: Unrecognized command 
<LF>? <CR><ETX> 

2.5) Output status bit meaning: 
Table5: The status bits definition:   

Bit 

Byte 1 (H

1

Byte 2 (H

2

Byte 3 (H

3

)  

0

 

0=stable 

0= not under capacity  01=normal work mode 

10= hold work mode       
00=not define          
11= not define   

1= not stable 

1= under capacity 

1

 

0= not at zero point  0= not over capacity 

1= at zero point 

1= over capacity 

2

 

always 0 

always 0 

0= gross weight 

1= net weight 

3

 

0= eeprom OK   

always 0 

always 0

 

1= eeprom error 

4

 

always 1 

always 1 

always 1 

5

 

always 1 

always 1 

always 1 

6

 

always 0 

always 1 

always 0 

7

 

parity Parity  parity 

 

21 

.

 

Relationship of capacity and the setting of P7, P8 and P9: 

Table6: (Capacity unit is kg or lb) 

Resolution 
set by P7 

Division set by P8 and P9

P8=0

 

0.0001 0.001 0.01  0.1  1  10 

500 0.0500 

0.500 

5.00 

50.0 

500 

5000 

600 

0.0600 

0.600 

6.00  

60.0  

600 

6000 

750 0.0750 

0.750 

7.50 

75.0 

750 

7500 

800 0.0800 

0.800 

8.00 

80.0 

800 

8000 

1000 

0.1000  1.000 10.00 100.0 1000 10000 

1200 

0.1200  1.200 12.00 120.0 1200 12000 

1500 

0.1500  1.500 15.00 150.0 1500 15000 

2000 

0.2000  2.000 20.00 200.0 2000 20000 

2400 

0.2400  2.400 24.00 240.0 2400 24000 

2500 

0.2500  2.500 25.00 250.0 2500 25000 

3000 

0.30000 3.000 30.00 300.0 3000 30000 

3500 

0.3500  3.500 35.00 350.0 3500 35000 

4000 

0.4000  4.000 40.00 400.0 4000 40000 

5000 

0.5000  5.000 50.00 500.0 5000 50000 

6000 

0.6000  6.000 60.00 600.0 6000 60000 

7000 

0.7000  7.000 70.00 700.0 7000 70000 

7500 

0.7500  7.500 75.00 750.0 7500 75000 

8000 

0.8000  8.000 80.00 800.0 8000 80000 

10000 

1.0000  10.000 100.00 1000.0 10000 100000 

12000 

1.2000  12.000 120.00 1200.0 12000 120000 

15000 

1.5000  15.000 150.00 1500.0 15000 150000 

20000 

2.0000  20.000 200.00 2000.0 20000 200000 

25000 

2.5000  25.000 250.00 2500.0 25000 250000 

30000 

3.0000  30.000 300.00 3000.0 30000 300000 

35000 

3.5000  35.000 350.00 3500.0 35000 350000 

40000 

4.0000  40.000 400.00 4000.0 40000 400000 

50000 

5.0000  50.000 500.00 5000.0 50000 500000 

60000 

6.0000  60.000 600.00 6000.0 60000 600000 

70000 

7.0000  70.000 700.00 7000.0 70000 700000 

75000 

7.5000  75.000 750.00 7500.0 75000 750000 

80000 

8.0000  80.000 800.00 8000.0 80000 800000 

100000 10.0000 100.000 

1000.00 

10000.0 

100000 

1000000 

 
 
 

Содержание MS140/300

Страница 1: ...MS140 300 Operation Manual Contents subject to change without notice Version 2 1 2012 01 10...

Страница 2: ...aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage radio lectrique edict par le minist re des Communications du Canada Warnings Electrical installation For your pro...

Страница 3: ...c device These devices emit RF and can cause unstable scale readings If your scale ever performs poorly try moving the scale to a different room or location Avoid using in condition of heavy vibration...

Страница 4: ...tivity 0 3mV V 3mV V must be fit to 0 2uV display grad 2 Input Resistor 80 3 Output Resistor 10 K LOADCELL WIRING PIN 1 RED EXCITATION PIN2 BLACK EXCITATION PIN3 GREEN SIGNAL PIN4 WHITE SIGNAL 3 KEY F...

Страница 5: ...ormal weighing mode and hold down time longer than 3s 1 Same meaning with that in 3 1 2 Same meaning with that in 3 1 3 Same meaning with that in 3 1 4 Power off the scale UNIT TARE ZERO HOLD PRINT UN...

Страница 6: ...oint dot in calibration unit P9 and capacity s unit in calibration P10 If the setting of FS is more than 199999 regardless of decimal point and weight unit the FS capacity will be shown by first four...

Страница 7: ...d data will blink If the indicator gets reasonable data it will calculate and store all parameters in EEPROM And then it will auto reset and display all segments of LCD full capacity like power on aga...

Страница 8: ...Detailed setup 5 1 P1 xy Auto off time xy 00 15 xy 00 means no auto off function xy 01 15 means the scale turns off to save power after 01 15 minutes when no variety of weight on scale and no key ope...

Страница 9: ...put compatible 8 data to the NCI SP1 format 5 5 P5 x Baud rate of RS232 communication x 0 1200bps x 1 2400bps x 2 4800bps x 3 9600bps x 4 19200bps 5 6 P6 x RS232 communication data format x 0 8N1 8 di...

Страница 10: ...lb Not available 0 0002 lb 0 0001 kg 0 0002 lb Not available 0 002 lb 0 001 kg 0 002 lb Not available 0 02 lb 0 01 kg 0 02 lb 0 5 oz 15 0 2 lb 0 1 kg 0 2 lb 5 oz 2 lb 1 kg 2 lb Not available 20 lb 10...

Страница 11: ...power off zero point as power on zero point and power off tare weight as current tare weight 5 16 P16 x Zero tracking range x 0 0d means no tracking x 1 0 25d x 2 0 5d x 3 1d x 4 1 5d x 5 2d x 6 3d x...

Страница 12: ...f the zero point is error it will return nine _ characters b The character will be for negative weight or a space character for positive 19 weight Minus sign follows after the first digit c Useless le...

Страница 13: ...ip of capacity and the setting of P7 P8 and P9 Table6 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 0 0 0001 0 001 0 01 0 1 1 10 500 0 0500 0 500 5 00 50 0 500 5000 600 0...

Страница 14: ...160000 1600000 100000 20 0000 200 000 2000 00 20000 0 200000 2000000 23 Table8 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 2 0 0005 0 005 0 05 0 5 5 50 500 0 2500 2 50...

Страница 15: ...as following drawings 1 Placed vertically 2 Placed horizontally 25 Key Definition summary KEY MODE DEFINITION HOLD PRINT Normal weighing mode Enter or exit HOLD mode output the data as per P4 P5 P6 s...

Страница 16: ...x 0 1 2 3 4 3 Baud rate for RS232 0 1200bps 1 2400bps 2 4800bps 3 9600bps 4 19200bps 27 P6 x 0 1 2 0 RS232 format 0 8N1 1 7O1 2 7E1 P7 xy 00 31 13 Resolution select 500 600 750 800 1000 1200 1500 200...

Страница 17: ...play 0 P16 x 0 1 2 3 4 5 6 7 8 8 Zero tracking range 0 0d no tracking 1 0 25d 2 0 5d 3 1d 4 1 5d 5 2d 6 3d 7 4d 8 5d P17 x 0 1 2 3 2 Data filter intensity 0 very weak 1 weak 2 middle 3 strong P18 x 0...

Страница 18: ...845 246 6717 Fax 44 0 845 246 6718 Email sales brecknellscales co uk Web site www brecknellscales co uk USA Brecknell 1000 Armstrong Drive Fairmont MN 56031 Toll Free 800 637 0529 Tel 507 238 8702 Fa...

Отзывы: