7.3.3
High Breath Rate Alarm
Property
Description
Alarm text
High Breath Rate
Priority
Medium
Alarm condition
A High Breath Rate alarm will be given when the alarm limit
has been exceeded for 15 seconds.
Possible cause
•
Mismatch between the Breath Rate setting and the alarm
setting.
•
Increased Breath Rate.
•
Too sensitive setting of the inspiratory trigger setting.
Reset criteria
The breath rate goes below the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
10 bpm to 50 bpm
•
Off
Setting resolution
1 bpm.
7.3.4
Low Breath Rate Alarm
Property
Description
Alarm text
Low Breath Rate
Priority
High
Alarm condition
A Low Breath Rate alarm will be given when the delivered
total breath rate is below the alarm limit for 15 seconds.
Possible cause
•
Mismatch between the Breath Rate setting and the alarm
setting.
•
The patient cannot trigger breaths because the inspiratory
trigger setting is too high.
•
Decrease in the patient
’
s spontaneous breathing.
•
Circuit disconnection.
Reset criteria
The breath rate goes above the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
•
4 bpm to 30 bpm.
•
Off
Setting resolution
1 bpm.
Alarms 97
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual