Supplemental Oxygen Supply Requirements
Property
Requirement
Maximum flow
30 l/min.
Maximum pressure
100 kPa
Supply source
Source equipped with rotameter. Exam-
ples of supply sources:
•
Oxygen cylinder
•
Central oxygen supply
•
Oxygen concentrator
Connector
The oxygen source shall be equipped
with the Breas Low Pressure Oxygen
Adapter, art. no. 005032
Connect the Oxygen supply
1
Connect the oxygen adapter (article no. 005032) to the oxygen supply
’
s tube.
2
Connect the oxygen adapter to the oxygen port at the back of the Vivo 3. See 3.3
, page 35 for detailed information.
3
If using a device for monitoring the oxygen saturation, connect it according to the
supplier instructions.
6.7.5
Using the Nurse Call Connection
The Vivo 3 can be connected to a nurse call system through a port on the accessory box.
When connected, alarms from the Vivo 3 will be forwarded to the nurse call system.
1
Connect the Accessory box to the Communication port at the back of the Vivo 3.
2
Connect the nurse call cable to the Nurse call/ Remote alarm port on the accessory
box.
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual