Property
Description
Alarm text
SpO2 Disconnected
Priority
Medium
Alarm condition
An SpO
2
Sensor Failure/Disconnection alarm will be given
if one of the conditions below appears:
•
An error signal is received from the sensor
•
No signal at all from the sensor is received within 2
seconds.
Possible cause
Faulty or disconnected sensor.
Reset criteria
Normal communication with the sensor is reestablished.
An information message remains until acknowledged by the
user.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
7.4.4
SpO
2
Artifact
This alarm requires a connected SpO
2
sensor.
Property
Description
Alarm text
Poor SpO2 Signal
Priority
Medium
Alarm condition
A poor SpO
2
Signal alarm will be given if perfusion is too
low or artifacts are detected by the sensor.
Possible cause
Check the sensor and its placement on the patient.
Reset criteria
An OK signal is received from the sensor or the sensor is
disconnected.
An information message remains until acknowledged by the
user.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Alarms 105
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual