7.3.14
High Pulse Rate Alarm
Property
Description
Alarm text
High Pulse Rate
Priority
Medium
Alarm condition
A High Pulse Rate alarm will be given when the measured
pulse rate exceeds the alarm limit for 15 seconds.
Possible cause
•
Insufficient ventilatory support.
•
Too low flow of bleed-in oxygen.
•
The EPAP value is set too high.
•
Bad positioning of the finger probe.
Reset criteria
The pulse rate goes back below the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
20 to 250 bpm (beats per minute)
Off
Setting resolution
5 bpm (beats per minute)
This alarm requires a connected SpO
2
sensor.
7.3.15
Low Pulse Rate Alarm
Property
Description
Alarm text
Low Pulse Rate
Priority
High
Alarm condition
A low pulse rate alarm will be given when the measured pulse
rate goes below the alarm limit for 15 seconds.
Possible cause
•
Bad positioning of the finger probe.
•
Too low flow of bleed-in oxygen.
•
Insufficient ventilatory support.
Reset criteria
The pulse rate goes back above the alarm limit.
Ventilator action
The Vivo 3 will continue treatment according to the current
settings.
Setting range
20 to 250 bpm (beats per minute)
Off
Setting resolution
5 bpm (beats per minute)
This alarm requires a connected SpO
2
sensor.
Alarms 103
Doc. 007232 Y-2
Vivo 3 Clinician
’
s Manual