Maintenance Instructions 16
2210.2657 Rev 3.0
NIPPY Clearway 2 Service Manual
.
Clean the valve assembly
Every 5 years remove the valve assembly and clean as described in chapter 7
of this manual.
Electrical Safety
Clearway 2 is a Class II, type BF medical electrical device.
Electrical equipment with dual isolation and body floating (isolated) applied
parts according to IEC 60601-1.
Regularly Inspect the AC mains cable and plug (part number: 2210.2701)
Electrical safety testing shall be carried out by persons experienced in the
electrical safety testing of medical devices and performed in accordance with
IEC 60601-1 or IEC 62353. A medical electrical safety tester shall be used to
test the Clearway 2.
It is recommended to carry out an electrical safety test at least
every 12 months
It is recommended to test:
•
Insulation Resistance
•
Leakage Currents:
o
Enclosure leakage current (touch current)
o
Patient applied part leakage current
Содержание Clearway 2
Страница 1: ...Service Manual Technical Instructions for Use 2210 2657 Revision 3 0...
Страница 25: ...25 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB...
Страница 26: ...Part identification 26 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Inside Rear Case...
Страница 27: ...27 Part identification NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Power PCB...
Страница 30: ...Functional Diagrams 30 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Control PCB block diagram...
Страница 31: ...31 Functional Diagrams NIPPY Clearway 2 Service Manual 2210 2657 Rev 3 0 Display PCB block diagram...
Страница 32: ...Functional Diagrams 32 2210 2657 Rev 3 0 NIPPY Clearway 2 Service Manual Pneumatic diagrams...