35
SPECIFICATIONS
Supply Voltage
-
100 - 240 V alternating current
Supply Frequency
-
47 - 63 Hz
Maximum Input Current
-
0.40
– 1.0 Amperes
Fuse Ratings
-
2 x T 1.6 A 20mm
Dimensions (mm)
Length
-
297 Width
- 223 Height
- 132
Weight
-
4.5 kg
Ambient Operating Temperature
-
40
o
C, 104
o
F Max
Digital Output
-
RS232 Isolated to 1500 Volts
All displayed readings expressed as
-
ATPD
Max. Output Pressure
-
38cm H2O(44cm fault condition)
Calibrated pressure Range
-
0 - 38cm H2O
Accuracy of pressure reading
-
+/- 3.0% F.S. +/-1% zero
Max. Output Flow
-
200 L/min. (unrestricted)
Max Volume Reading
-
2000 millilitres
Accuracy of volume reading
-
Estimated
Accuracy of Flow reading
-
+/-10%
Inspiratory Trigger
-
0.14
– 2.21 L/sec
2
Expiratory Trigger
-
0.28
– 1.67 L/sec
2
Low Flow Alarm
-
0
– 200 lpm
High Flow Alarm
-
0
– 200 lpm
Inspiratory Time
-
0.7
– 3.0 seconds
Back-up Rate
-
6 - 43 Breaths per minute
Type of protection against electric shock
-
Class 1 equipment
Degree of protection against electric shock
-
Type B to EN 60601-1
Mode of operation
-
Continuous
IP rating
-
X0
Storage environment
-
-20 to 50
O
C
5
– 85% RH
260
– 1100 mBar atmospheric pressure
Internal battery
-
18.75Vdc 116Whr
Running time
-
4 -12 hours depending on settings and leak
External battery
-
18.75Vdc 116Whr
Running time
-
4 -12 hours depending on settings and leak
Protection against flammable anaesthetic mixtures
- Not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE
International Standards
BS EN60601-1 1990, EN 10651- 6 2004
Safety of Electromedical Instruments, General Requirements
Electromagnetic Compatibility (In accordance with the EMC Directive 89/336/EMC)
Breas Medical Ltd declares that the NIPPY 3+ Ventilator complies with the following EMC standards.
EN60601-1-2: 2001
Test results available for review from Breas Medical Ltd
0086