The tone

Each generated pulse produces a 

tick

. When a stimulation current

flows to the electrodes (greater than 2% of the maximum stimula-
tion adjustable), the 

tick

 becomes louder.

If the stimulation current flowing exceeds the current set, then
taking into account the measuring tolerance (see page 12 - Techni-
cal data - Measuring tolerance), the Stimuplex

®

 HNS 11 institutes

an error message consisting of several tone groups of rapid
frequency (approx. 6 Hz). The device is no longer operational.

Each time a button is pressed a short tone is produced of a differ-
ent frequency as audible confirmation.

click

 confirms that the instrument has been switched on.

Initial set up, maintenance, and general advice

Testing the Stimuplex

®

 HNS 11 before an application

Switch on the Stimuplex

®

 HNS 11 nerve stimulator by turning the

control knob clockwise from position 

O

. A short click confirms that

switching on has taken place. When the self-test has been com-
pleted, the unit changes over immediately to the normal operating
mode. You can now select the frequency, pulse width and current
range according to the application required (see Section ”Operat-
ing the nerve stimulator”, p. 7).

With the set current indication (mA  ), do not connect both elec-
trodes - check both current ranges (1 mA max and 5 mA max) up
to maximum.
Switch to actual current indication (mA

) - both mA ranges give

a should reading of + 0.00 mA, independet from the set current.

Connect the two electrodes (clip and plug) together: check the mA
value in each case up to the maximum. All mA values have a ”

+

sign in front of the display. As soon as current flows, the LED above
the display flashes in time with the stimulation current. The 

~

symbol appears next to the 

 and the 

tick

 becomes louder.

Set an actual current of approx. 0.30 mA and switch to set current
(mA  ): the two values should be the same (see Page 12 - Techni-
cal data - Measuring tolerance).

If all adjustments have been found to be correct, your Stimuplex

®

HNS 11 is ready for application.

11

Содержание Stimuplex HNS 11

Страница 1: ...Stimuplex HNS 11 Nerve Stimulator for Regional Anaesthesia Regional Analgesia Neurology User manual...

Страница 2: ...Distributor B Braun Melsungen AG P O Box 11 20 D 34209 Melsungen Tel 0 56 61 71 0 Fax 0 56 61 71 3811 Manufacturer Stockert GmbH B tzinger Stra e 72 D 79111 Freiburg Germany...

Страница 3: ...D display 10 The tone 11 Initial set up maintenance and general advice 11 Testing the Stimuplex HNS 11 before an application 11 Special technical features 12 Technical data 12 Battery 13 Cleaning and...

Страница 4: ...the self test is completed the stimulator immediately switches to its normal operating mode The LEDs on the button panel indicate the operating conditions selected To conserve power these extinguish a...

Страница 5: ...equence A sign appears in front of each measurement displayed When a current is set at O the orange LED starts to blink at 0 02 mA when 1 mA max is set and at 0 10 mA when 5 mA max is set 4 When the s...

Страница 6: ...lows precise adjustments for extremely thin and technically sophisticated needles The stimulation frequency and the pulse width can be varied for different applications The Stimuplex HNS 11 nerve stim...

Страница 7: ...in immediately when a button is pressed but this does not change the operating mode The operating mode can only be changed when the displays are lit To switch off the nerve stimulator turn the control...

Страница 8: ...e LCD display Pressing this button does not affect the Stimu plex HNS 11 functions All settings remain unchanged Selecting the current display mode During operation of the nerve stimulator the stimula...

Страница 9: ...rs in front of the current displayed Current adjustment range Setting range from 0 00 to 5 00 mA Use the controller to dial a stimulation current up to 5 mA The current is displayed in 0 02 mA digits...

Страница 10: ...s next to the symbol in the same stimulation frequence when the stimulation current is flowing and the ticking sound gets louder When the actual value display is selected the LCD display additionally...

Страница 11: ...ing on has taken place When the self test has been com pleted the unit changes over immediately to the normal operating mode You can now select the frequency pulse width and current range according to...

Страница 12: ...technology has made it possible to integrate performance characteristics which until now could only be provided in large systems For example selection of cur rent display mode set and actual value di...

Страница 13: ...compartment is in the base of the nerve stimulator Open the compartment by lightly pressing the OPEN symbol At the same time slide the cover in the direction of the arrow Change the battery ensuring...

Страница 14: ...se can be used for disinfection Caution The following substances must not be used trichloroethylene acetone methyl ethyl ketone benzene metha nol cellulose thinner 1 propanol and 2 propanol Maintenanc...

Страница 15: ...For usage of HNS 11 please refer to the individual local law restricts eg maintaining a logbook etc In case of conducting safety checks on the device please refer to the enclosed inspection sheet 15...

Страница 16: ...he cables must not exhibit any damage Only original accessories from B Braun labelled with the CE sign should be used Regarding the disposal of the nerve stimulator and its accessories you should refe...

Страница 17: ...refer to p 21 points 8 and 9 If the battery has leaked the Stimuplex HNS 11 must not continue to be operated for safety reasons It must be returned to the manufacturer for correct cleaning The same ap...

Страница 18: ...ex D Stimuplex A and Contiplex D needles Knob for sterile handling 4892283 1 Electrode cable for Stimuplex D 4892070 1 Stimuplex A and Contiplex D needles Connection cable for Contiplex A needles 4892...

Страница 19: ...dle D 28 055 C 18 G 15 bevel 1 3 x 55 mm 4894235 10 needle D 18 055 C 18 G 30 bevel 1 3 x 55 mm 4894243 10 with polyamide catheter 0 45 x 0 85 x 1000 mm needle D 28 110 C 18 G 15 bevel 1 3 x 55 mm 489...

Страница 20: ...lized with short pulses 0 1 ms without triggering pain sensations For pure sensory nerves a longer pulse 0 3 or 1 0 ms is recommended When using unipolar needles needles with an insulated needle shaft...

Страница 21: ...of the stimulator the symbol appears in the display in time with the stimulation pulse to indicate that it is ready to operate Caution A change to the stored parameters can only be made when the LEDs...

Страница 22: ...o be achieved depends on the pulse width and the geometry of the conductive tip of the needle Warning When using Stimuplex or Contiplex needles it must be noted that muscular contractions with a curre...

Страница 23: ...naesthesie 13 143 6 Kaiser H Niesel HC Hans V Klimpel L 1990 Untersuchungen zur Funktion peripherer Nervenstimulation f r die Durchf h rung von Nerven und Plexusblockaden Regional Anaesthesie 13 172 7...

Страница 24: ...0 mA 0 1 3 00 mA 0 15 4 00 mA 4 00 mA 0 1 4 00 mA 0 15 5 00 mA 5 00 mA 0 1 5 00 mA 0 15 24 Appendix B Conversion table Amplitude mA and pulse width ms to charge nano Coulomb nC Amplitude Pulse width C...

Страница 25: ...Stimulator Max stimulation current as a function of tissue resistance Max stimulation current mA Battery voltage V Stimuplex HNS 11 Nerve Stimulator Operating range Max stimulation voltage V Battery...

Страница 26: ...lse width and minimum 0 2 mA with 0 1 ms pulse width allow reliable location of nerves Fully insulated needle with bare tip Stimuplex D Contiplex D 26 Distance from the nerve mm Bare and insulated nee...

Страница 27: ......

Страница 28: ...n AG P O Box 1120 D 34209 Melsungen Tel 0 56 61 71 0 Fax 0 56 61 71 3811 www bbraun com W 01 08 01 7 Nr 604 3895 hospital care Caution Federal U S A law restricts this device to sale by or on the orde...

Отзывы: