Unit Elements
4
Perfusor® compact, 2.1 gb
4 - 3
Open unit
1. Loosen 5 screws from the bottom.
2. Open housing carefully, then
3. Pull off the ribbon cable from the E-Module and the
connection cable from the motor. Hold the white board
holder on the E-Module when disconnecting!
4. Dismount both housing halves.
Always check A-Module before replacing the board.
Other modules can only be exchanged without danger of
consequential damage if there is no overvoltage.
Connect mains power cable when the housing is open. Measure
voltage parallel with capacitor C3. The set value is 6.2 to 6.8 volt.
Fig.: 4 - 1
Ribbon cable to drive
board
Motor cable
Fig.: 4 - 2 A-Module board (up to serial number 38100)
2.1
Содержание perfusor compact
Страница 1: ...Perfusor compact Service Manual 0 10 3 1 0 Version 3 1 English...
Страница 10: ...Contact Persons 0 0 10 Perfusor compact 2 1 gb For your notes 2 1...
Страница 14: ...System Overview 1 1 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 20: ...Software 2 2 6 Perfusor compact 2 1 gb For your notes 2 1...
Страница 56: ...Unit Elements 4 4 18 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 62: ...Checks after Repair 5 5 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 64: ...Maintenance 6 6 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 72: ...Procedural Instructions on the TSC 8 8 6 Perfusor compact 2 1 gb 2 1 For your notes...
Страница 74: ...Test Equipment and Special Tools 9 9 2 Perfusor compact 2 1 gb For your notes 2 1...
Страница 78: ...Spare Parts List 10 10 4 Perfusor compact 2 1 gb For your notes 2 1...
Страница 80: ...Revision Documentation 11 11 2 Perfusor compact 2 1 gb 2 1...
Страница 85: ......