12
Product specifications
Electrical characteristics:
2 X AA (1,5 Vcc c/u) 3.0 Vcc
Type:
No touch + forehead thermometer (NTF3000LA/NTF3000LAD1)
Measuring range:
34.4 °C – 42.2 °C (93.9 °F – 108 °F)
Resolution:
0.1 °C (0.1 °F)
Laboratory Accuracy:
±0.2 °C for the range 35 °C – 42 °C (±0.4 °F for 95 °F – 107.6 °F)
±0.3 °C (outside this temperature range)
(Ambient Temperature: 15 °C to 40 °C (59 °F to 104 °F))
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic: Audio:
Normal temp range = single long beep for 2 sec duration
Fever = Red or Yellow temp range: 10 short beeps
Operating temperature:
15 °C – 40 °C (59 °F to 104 °F)
Automatic Switch-off:
Approx. 60 seconds after last measurement has been taken
Weight:
99.5 g (with batteries), 77.1 g (w/o batteries)
Long term storage ranges
Storage/transport temperature:
-25 °C to 60 °C (-13 °F – 140 °F)
Humidity:
15–95% non-condensing
Battery:
(2) AA Batteries - at least 1000 measurements
Pressure:
700-1060 hPA (0,7-1,06 atm)
This product conforms with the provisions of the European directive 93/42/EEC (Medical Device
Directive). This infrared thermometer meets requirements established in ASTM Standard E 1965-98 for the
thermometer system. Full responsibility for the conformance of the product to the standard is assumed by
Kaz USA, Inc., 250 Turnpike Rd., Southborough, MA 01772, USA.
ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C
(96.8 °F to 102.2 °F) for infrared thermometers is ±0.3 °C (±0.5 °F), whereas for mercury-in-glass and
electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).
This device conforms to the following standards:
EN 60601-1:2006 Medical electrical equipment. General requirements for basic safety and
essential performance
EN 80601-2-56 Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
ASTM 1965E – ASTM Standard for Infrared Clinical Thermometers
EN 12470-5: 2003 «Clinical thermometers» – Part 5: Performance of infrared ear thermometers (with
maximum device)
EN 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential
performance – Collateral Standard: Electromagnetic compatibility – Requirement and tests