57
Recharging time Approx. 6 hours
Weight Approx. 1.4 kg
Dimensions (W x H x D) 214 x 68 x 124 mm
Volume preselection 0.1 - 99.99 ml in increments of 0.01 ml
100.0 – 999.0 ml in increments 0.1 ml
1000 – 99999 ml in increments 1 ml
Time preselection 00:01 – 99:59 h
Accuracy of set delivery rate typ. ± 5 % at rate 100 ml/h,
temperature 22°C, measurement time 24h
Max. Volume in case of single fault
condition For incorrect dosages of 1,4 ml due to
malfunctions of the device the pump will
automatically shut off
Technical inspection (safety check) Every 2 years
Administration Set Change Interval Pumping accuracy is maintained for a
minimum of 24 hours.
Multiple lines connected to one Connecting multiple infusion lines with
patient port different flow rates may affect the rate
for all infusions past the point of
connection.
Rate increments 0.1 - 99.99 ml/h in increments of
0.01 ml/h
100.0 - 999.9 ml/h in increments of
0.1 ml/h
1000.0 - 1200 ml/h in increments of
1 ml/h
Accuracy of bolus infusion typ. ± 5 % as of a bolus volume > 1 ml
KVO-rate Delivery rate
≥
10 ml/h: KVO-rate 3 ml/h
Delivery rate < 10 ml/h: KVO-rate 1 ml/h
Delivery rate < 1 ml/h: KVO-rate =
set rate (default setting 0.1 ml/h)
Computer connection USB connection in combination with
B. Braun interface lead CAN SP
(8713230) including electrical
insulation. Please pay attention to
safety notices.
Air detector Technical sensitivity:
Detection of air bubbles
≥
0.01 ml
Alarm triggering:
Single bubble alarm: 0.02 – 0.3 ml
(default 0.3 ml)
Cumulative air alarm: 0.5 – 3.8 ml/h
(default 1.5 ml/h)
Resolution: 0.01 ml
Chapter 9
TECHNICAL DATA