Product specification
Type:
No touch 3 in 1 thermometer (BNT100)
Measuring range:
Body: 34 .0 °C – 43 .0 °C (93 .2 °F – 109 .4 °F)
Food/Bath: 0 °C – 100 .0 °C (32 .0 °F – 212 .0 °F)
Measuring location:
Forehead
Resolution:
0 .1 °C (0 .1 °F)
Laboratory accuracy
- Body temperature mode:
±0 .2 °C for the range 35 .0 °C – 42 .0 °C
(±0 .4 °F for 95 .0 °F – 107 .6 °F)
±0 .3 °C (±0 .5 °F) for the range outside
35 .0 °C – 42 .0 °C (95 .0 °F – 107 .6 °F)
- Food/bath temperature mode: ±2 .0 °C for the range 0 .0 °C – 100 .0 °C
(non-medical use)
(±3 .6 °F for 32 .0 °F – 212 .0 °F)
Clinical accuracy:
Clinical performance and protocols are available upon request
This thermometer displays a calculated oral equivalent estimate
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
Audio:
Normal temp range = Green temp range: 1 long beep for 0 .4
second duration
Fever = Yellow temp range: 1 long beep for 0 .4 second duration
High fever = Red: 10 short beeps (0 .02 seconds on, 0 .13 seconds off)
Operating temperature:
15 °C – 40 °C (59 .0 °F to 104 °F)
Automatic Switch-off:
Within approx . 20 seconds
Weight:
93 .2g (w/ batteries) and 71 .5 (w/o batteries)
Size:
55mm x 38mm x 149mm
Long term storage ranges
Storage/transport temperature:
-25 °C to 55 °C (-13 °F to 131 °F )
Humidity:
15–95% non-condensing
Battery:
(2) AAA Batteries - at least 500 measurements
Pressure:
700-1060 hPA (0 .7-1 .06 atm)
Memory size:
10
The expected service life of the thermometer is 5 years .
If necessary, the circuit diagram and component list can be provided for Authorized
Technician Maintenance .
The BNT100US includes the following items: No touch thermometer,
2 AAA batteries, owner’s manual and quick reference guide .
This infrared thermometer meets requirements established in ASTM Standard
E 1965-98 (for the thermometer system) . Full responsibility for the conformance of
the product to the standard is assumed by Kaz USA, Inc ., a Helen of Troy Company,
Health & Home, 400 Donald Lynch Blvd ., Suite 300, Marlborough, MA 01752 .
ASTM laboratory accuracy requirements in the display range of 37 to 39 °C
(98 to 102 °F) for IR thermometers is +/- 0 .2 °C (+/- 0 .4 °F), whereas for mercury-in-glass
and electronic thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is +/- 0 .1 °C (+/- 0 .2 °F) .