Braun AESCULAP 3D EinsteinVision 3.0 Скачать руководство пользователя страница 3

 

AESCULAP

®

 3D 

Einstein

Vision

® 3.0 

Flow 50 Insufflator

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On start-up - First screen offers surgical specialties 

 

Select appropriate 

Indication

” 

- f

ollowed by ”Veress” or “High Flow” 

to move to main screen.  

(Cardiac mode has 2 functions - 

“Vessel Harvesting” or “Field Flooding”

 (Field flooding is flow only)

 

 

 

 

 

 

 

 

Pre-set pressure can be set up to 30 mmHg 

Nb. 15 mmHg is set as the Safety Threshold “

Safety Threshold

” will flash repeatedly

 

Press 

+

 when the safety threshold message has gone off, to take pressure beyond 15 mmHg 

 

Press and hold the Flow 

+

  or 

 

buttons to go between “Low Flow”, Medium Flow” and “High Flow”

 

Low Flow = 3 L/min    Medium Flow = 20 L/min    High Flow = 40 L/min 

 

Insufflation is started and stopped using the “Start” button (Bottom Right)

 

(When insufflati

on is active, button displays “Stop”.  Insufflation is stopped by pressing “Stop”

 

 

CO

2

 volume delivered is reset by pressing the reset button 

Optional Smoke 

Evacuation Filter 

Housing 

Active Smoke Evacuation 

Actual 

Actual 

Volume 
& Reset 

System Settings 

Insufflation 

Gas Outlet 

Heated Gas 

Element 

(Optional) 

Start / Stop 
Insufflation 

Smoke Evacuation 

Footswitch 

(Optional) 

Pressure & Flow  

Pre-sets 

On / Off 

Содержание AESCULAP 3D EinsteinVision 3.0

Страница 1: ...on 3 0 Quick System Guide 32 3D Monitor 21 2D Monitor Keyboard 50L Insufflator LED Light Source 3D 2D CCU Digital Documentation 3D Scope Camera Holder 2D Camera Holder CO2 Gas Cylinder Storage A4 Ther...

Страница 2: ...ht side of 21 monitor At the bottom of the monitor 4 Plug in integrated 3D light lead 5 if using 3D scope Plug in 3D camera head 6 if using 3D scope Set Insufflator to desired pressure 7 and flow rate...

Страница 3: ...ne off to take pressure beyond 15 mmHg Press and hold the Flow or buttons to go between Low Flow Medium Flow and High Flow Low Flow 3 L min Medium Flow 20 L min High Flow 40 L min Insufflation is star...

Страница 4: ...irmly into light aperture Do not rotate cable release Rotate clockwise to increase light output Press light source control to activate Deactivate light output Rotate anti clockwise to decrease light o...

Страница 5: ...ds Camera White Balance can be performed from camera head remote button Rt or from CCU WB button Camera head button Long press white button 2sec CCU WB Button Short press 2sec Long press of CCU white...

Страница 6: ...AESCULAP 3D EinsteinVision 3 0 2D 3DCamera Heads 3D Integrated Camera Head Button Configuration 2D Camera Head Button Configuration...

Страница 7: ...MAKE ANY CHANGES TO THE MONITOR SETTINGS There is a 3D 2D button on the monitor DO NOT USE THIS BUTTON to swap between 3D and 2D Using the 3D 2D button on the monitor to select 2D effectively renders...

Отзывы: