Appendix- Electromagnetic declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The EndoSync (hereafter the TR-CM) is intended for use in the electromagnetic environment specified below. The customer
or the user of the TR-CM should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TR-CM uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The TR-CM is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions*1
IEC61000-3-2
Class A
Voltage fluctuations/ flicker emissions*1
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The TR-CM is intended for use in the electromagnetic environment specified below. The customer or the user of the TR-CM
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±2, 4, 6 kV contact
±2, 4, 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least 30
%.
Electrical fast tran-
sients/bursts
IEC 61000-4-4
±2 kV for power supply
lines ±1 kV for input/
output lines
±2.0 kV for power sup-
ply lines
*2
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±0.5, 1, 2 kV line(s)
to line(s) ±0.5, 1 kV
line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in
UT ) for 5 cycles
70% UT (30% dip in
UT ) for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec
0% UT (>95% dip in
UT) /0.5 periods
40% UT (60% dip in
UT ) /5 periods
70% UT (30% dip in
UT ) /25 periods
0% UT /5 sec.
Mains power quality should be that of a typical
commercial or hospital environment. If user of
the TR-CM requires continued operation during
power mains interruptions, it is recommended
that the TR-CM be powered from an uninter-
ruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3.15 A/m
Power frequency magnetic field should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.