Charging Bay Installation:

• Loosen screws and remove Charging Bay Cover.  
• Align front Charging Bay tab with Power Unit slot.  
• Lower the Charging Bay into the corresponding side

slots and power input connection. Ensure Charging Bay
tabs are aligned with Power Unit slots.

• Apply light force to seat fully.
• Tighten screws with screwdriver.
• Verify  that  the  Charging  Bay  is  flush  with  the  top

surface of the 4-Bay Power Unit and that the front
tab is engaged into power slot.

Charging Bay Removal:

• Loosen Screws.
• Remove Charging Bay.
• Install Charging Bay Cover (PM-X00-521).

The 4-Bay Power Unit can be configured with a combination of Charging Bay(s) and Bay Cover(s). The 4-Bay
Power Units (PM-X00-520 and PM-X00-522), Standard Battery Pack Charging Bay (PM-X00-731) and Charging
Bay Cover (PM-X00-521) are sold separately. The power unit charges up to four battery packs simultaneously.

Cautions:

• DO NOT over tighten screws.
• DO NOT connect power cord to the electrical connector inside the charging bay.

Tool Required:

• Standard Phillips #2 (medium size) Screwdriver.

Charging Bay – PM-X00-731

Installation & Removal

Power Input
Connection

Charging 
Bay Tab

Charging Bay Cover

Charging 
Bay Tab

Содержание BSP MAX II

Страница 1: ...PM X00 520 PM X00 522 PM X00 731 IR 12 13 SURGICAL POWER ACCESSORIES Rx Only ...

Страница 2: ...ation and Removal Page 8 4 Bay Power Unit PM X00 520 PM X00 522 Page 9 4 Bay Power Unit and Operation Page 9 Battery Pack Charging Page 10 Troubleshooting Page 11 Care and Maintenance Page 13 Cleaning Recommendations Page 14 Specifications Page 16 Repair Service Page 23 Warranty Page 24 Return Goods Policy Page 25 Product Disposal Page 26 Product Ordering Information Page 26 ...

Страница 3: ...and effective use care cleaning sterilization and long term maintenance of BSPMAX II Large Bone Power System Introduction The BSPMAX II 4 Bay Power Units PM X00 520 and PM X00 522 are units designed to accept up to four modular Standard Battery Pack Charging Bays PM X00 731 The modular standard battery pack charging bay is intended to charge BSPMAX II 9 6V Battery Pack PM X00 710 and 12V Battery P...

Страница 4: ... a damaged power cord or plug DO NOT modify the ground of the 4 Bay Power Unit power cord Install the power cord of the 4 Bay Power Unit directly into electrical outlet DO NOT disassemble or service the 4 Bay Power Unit or Charging Bay Return to Brasseler U S A Medical LLC for service or repair Failure to comply may result in electric shock or fire Always disconnect the power cord from the 4 Bay P...

Страница 5: ...rom Charging Bays when 4 Bay Power Unit is off to avoid battery discharge and or irreparable damage to battery cells Catalog number Caution Date of manufacture Do not discard Dispose of product or recycle in accordance with local laws and regulations Do not immerse h Y N Interference Manufacturer Serial Number Shock hazard type B Temperature limit M f l ...

Страница 6: ...r Cord Connection Charging Bay Modular battery charging bay for BSPMAX II battery packs Power Unit will accept up to four 4 Charging Bays Charging Bay Cover Shields Power Unit s cooling fan and Charging Bay connector when not in use Cover should only be removed when installing a Charging Bay Indicator Lights Provide battery status information when illuminated Charge Yellow Battery Pack is charging...

Страница 7: ... Removal Loosen Screws Remove Charging Bay Install Charging Bay Cover PM X00 521 The 4 Bay Power Unit can be configured with a combination of Charging Bay s and Bay Cover s The 4 Bay Power Units PM X00 520 and PM X00 522 Standard Battery Pack Charging Bay PM X00 731 and Charging Bay Cover PM X00 521 are sold separately The power unit charges up to four battery packs simultaneously Cautions DO NOT ...

Страница 8: ... fuses 5 Replace fuse holder 6 Connect the power cord 7 Set the main power switch to ON 4 Bay Power Units PM X00 520 110V PM X00 522 230V Power Cord Connection Extra female chaining connector DO NOT USE Main Power Switch 4 Bay Power Unit Operation 1 Set the main switch to the OFF position O 2 Connect the female end of the power cord to the power cord connection on the back of the power unit 3 Conn...

Страница 9: ...ed on Yellow pulsing Primary diagnostic test to determine voltage level Yellow solid Battery Pack is charging Green solid Battery Pack is charged and ready for use Red pulsing Secondary diagnostic test to determine charging capacity Red solid Charging cycle failed and battery pack should be disposed of properly refer to Battery IFU for Product Disposal Information Battery Pack Charging Charging Ba...

Страница 10: ...ays Charging bay is not installed correctly See installation and removal Power unit fuses are open blown Replace with appropriate rated fuses See 4 Bay Power Unit fuse replacement Battery pack is not seated properly Fully seat battery onto charging bay Charging contacts are dirty See cleaning recommendations Battery pack malfunctions Replace battery pack After inserting battery pack into charging ...

Страница 11: ...ecurring power unit open blown fuses Charging board circuit failed SendpowerunittoBrasseler U S A Medical LLC forservice Fuses have wrong rating Replace with appropriate rated fuses See 4 bay power unit fuse replacement A power surge occurred Install power surge suppressor Indicator light does not illuminate when battery pack is installed Charging bay malfunctioned SendchargingbaytoBrasseler U S A...

Страница 12: ...edical LLC Service Department for routine preventive maintenance every twelve 12 months Follow a regular care regimen that includes routine cleaning and a thorough inspection for damage Routine preventive maintenance performed every twelve 12 months by the Brasseler U S A Medical LLC Service Department can increase the reliability and extend the life span of your BSPMAX II Large Bone Power System ...

Страница 13: ...ay Power Unit or Charging Bays in liquid DO NOT use solvents lubricants or other chemicals to clean the 4 Bay Power Unit or Charging Bays unless otherwise directed DO NOT allow water to collect on the 4 Bay Power Unit or Charging Bays DO NOT sterilize the 4 Bay Power Unit or Charging Bays DO NOT clean the charging contacts with abrasives Prior to changing fuses set power unit main switch to OFF O ...

Страница 14: ... lightly dampened with a non abrasive hospital disinfectant 5 Dry immediately with a clean lint free soft cloth 6 Clean charging contacts using a cotton swab lightly with isopropyl alcohol DO NOT clean the charging contacts with abrasives 7 Ensure equipment is completely dry before reconnecting to power 8 Inspect for damage or malfunctioning Return damaged components to Brasseler U S A Medical LLC...

Страница 15: ...n Protection Class 1 Protection Class 1 Class of Rating Continuous operation Continuous operation Socket Outlet 5A 100 120V 50 60 Hz 3A 220 240V 50 60 Hz 1 Performance Size 17 x 8 5 8 x 6 1 4 Weight 16 75 lb 4 Bay Power Unit with 4 charging bays Size 7 1 2 x 3 1 2 x 4 Weight 1 8 lb Charging Bay Size 17 x 8 5 8 x 2 1 4 Weight 9 55 lb 4 Bay Power Unit 2 Physical Characteristics ...

Страница 16: ...ompliance EN60950 1 IEC60950 1 CSA 60950 1 UL 60950 1 Operating Ambient temperature 50 F to 70 F 10 C to 21 C Relative Humidity 30 75 Atmospheric Pressure 700hPa to 1060hPa Transport Ambient temperature 4 F to 158 F 20 C to 70 C Relative Humidity 10 100 Atmospheric Pressure 500hPa to 1060hPa 4 Environmental Requirements ...

Страница 17: ...dance RF emissions CISPR 11 Group 1 The BSPMAX II battery charger uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class A The BSPMAX II battery charger is suitable for use in all establishments including domestic establishments and those directly connected to the ...

Страница 18: ...ment should be used no closer to any part of the BSPMAX II Battery Charger including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 67 P d 1 67 P 80 MHz to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the trans...

Страница 19: ... covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 KV for power supply lines 1 KV for input output lines 2 KV for power supply lines 1 KV for input output lines Main power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 KV differential mode 2 KV common mode 1 KV differential mode ...

Страница 20: ...Reduction 10 ms 60 Reduction 100 ms 30 Reduction 500 ms 95 Reduction 5 sec Power frequency 50 60Hz Magnetic field IEC 61000 4 8 3 A m 3 A m At 50 Hz Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Guidance and Manufacturer s Declaration Electromagnetic Emissions Specifications continued 5 Electromagnetic Compa...

Страница 21: ...0 KHz to 80 MHz d P 80 MHz to 800 MHz d P 800 MHz to 2 5 GHz d P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 70 3 70 7 37 100 11 70 11 70 23 30 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rati...

Страница 22: ...air This warranty applies to the actual work performed Products must be decontaminated and sterilized before returning Note It is unlawful to ship contaminated non sterilized products Contact a Customer Service Representative at Brasseler USA Surgical Power Accessories at 877 834 7133 to request repair preventive maintenance or a loaner instrument If available loaner instruments will be supplied i...

Страница 23: ...ship for a period of ninety 90 days from the invoice date The warranty is limited to the repair of the product without charge This warranty is nontransferable In addition this warranty is void in the event of any of the following abuse misuse neglect or use in other than normal surgery environment disassembly alteration or unauthorized repair or in the event that the product has not been used in r...

Страница 24: ...ale Packaging and Shipping Returned items should be packaged in original packaging Credit will not be issued for items damaged in return shipment due to packaging inadequacy Note It is unlawful to ship contaminated non sterilized products Contact a Customer Service Representative at Brasseler USA Surgical Power Accessories at 800 535 6638 to obtain a Return Merchandise Authorization RMA number Ple...

Страница 25: ...30 1 4 IN KEYLESS CHUCK PM X08 935 3 MM KEYLESS CHUCK PM X08 940 ZHS ZIMMER HUDSON STRYKER UNIVERSAL ATTACHMENTS REAMERS PM X08 911 1 4 IN JACOBS HI TORQUE W KEY EXTENDED LENGTH PM X08 950 HUDSON EXTENDED LENGTH PM X08 955 ZIMMER EXTENDED LENGTH PM X08 960 AO EXTENDED LENGTH PM X08 965 ZHS ZIMMER HUDSON STRYKER UNIVERSAL STANDARD LENGTH PM X08 970 1 4 IN JACOBS HI TORQUE W KEY STANDARD LENGTH ATTA...

Страница 26: ...t affiliated with any of the above and makes no claim to copyrights or trademarks which are the property of these companies Brasseler USA and BSPMAX are trademarks of Peter Brasseler Holdings LLC or its affiliates All other trademarks are trademarks of their respective owners or holders Colors specifications and product availability subject to change Brasseler USA Surgical Power Accessories produc...

Отзывы: