1) Brush off the debris at the clean head holes. (Fig. 6) Do not use a wire brush.
2) Half fill a cup with clean water.
3) Rotate the handpiece and immerse half of the handpiece head. (Fig. 7)
4) Rotate and stop intermittently the handpiece about 2 to 3 seconds. (About 3 times)
6) Stop the handpiece and wipe it dry. If the dirt is not removed from the hole, remove
the turbine and clean it with a brush.
Fig. 6
Clean Head Hole
Clean Head Hole
Clean Head Hole
CAUTION
Use only clean water to maintain the Clean Head System.
Fig. 7
7. Clean Head System Maintenance