Operator Manual

 

 
 

 

 
 
 

Caution:  U.S. Federal law restricts this device to sale by 

or on the order of a physician

 

 
 

 

Содержание DL700

Страница 1: ...Braemar Inc DL700 DL700 DL700 DL700 Operator Manual Caution U S Federal law restricts this device to sale by or on the order of a physician ...

Страница 2: ...pt for the express warranties stated above Braemar disclaims all warranties including implied warranties of merchantability and fitness The stated express warranties are in lieu of all obligations of liabilities on the part of Braemar for damages including but not limited to special indirect or consequential arising out of or in connection with the use or performance of Braemar products Any action...

Страница 3: ... Needed 10 4 2 Recommended Steps 10 5 Operation 12 5 1 Materials Needed 12 5 2 To Start Recording 12 5 3 To Stop Recording 14 5 4 Analyzing the ECG Data 14 Service And Maintenance 15 6 Maintenance 16 6 1 Replacing the Clock Battery 16 7 Troubleshooting Guide 18 7 1 Problem Description and Recommended Solution 18 8 Service Items 20 Appendix A 21 9 Electrode Placement 7 Lead 22 10 Electrode Placemen...

Страница 4: ...ies and Receptacle The DL700 provides a battery compartment with secure receptacles for inserting two standard AA alkaline batteries A polarity guard feature is built in to prevent damage in case the batteries are installed incorrectly Cover Labeling The label on the outside of the cover provides information about the company supplying the recorder while the label on the inside of the cover provid...

Страница 5: ...ient to leave the clinic DL700 Holter Monitor 4 Gain Switch A gain switch is provided to allow the medical technician to adjust signal amplitude Three positions are provided 1 2X 1X and 2X Position indications are on the label above the switch Operating LED This is the green LED just to the right of the gain switch Fig 1 It provides current status information on the recorder See Operation Section ...

Страница 6: ... card interface Physical Dimensions 6 00 x 3 50 x 95 Weight 10 ounces with batteries Enclosure Molded plastic Operating Position Any orientation Operating Temperature 0 C to 45 C Non operating Temperature 40 C to 60 C Operating Humidity 10 to 90 non condensing Electrical Gain Settings 1 2X 1X and 2X Battery Two AA Alkaline disposable Connector 7 pin 3 diff 1 com single lead wires Test Jacks 3 chan...

Страница 7: ...the unit WILL NOT RECORD and the LED WILL NOT BLINK 3 2 Setting the Clock Time and Date on the digital clock can easily be set by the following procedure using the two small access holes labeled set and select The holes are located next to the Patient Event Button shown in Figure 1 a Insert and press a pointed object into the top access hole labeled set to start the setting process b Press the poi...

Страница 8: ...out good skin preparation both noise and artifact will frequently degrade the patient s ECG to the point of being unusable by the physician 4 1 Materials Needed a For 3 channel recording use seven 7 Silver Silver chloride disposable pre jelled electrodes designed for 24 hour Holter monitoring Do not use 12 lead ECG or Stress Test Electrodes Use electrodes designed for longer term 24 hour monitorin...

Страница 9: ...t s personal data on the Flash memory card label including the doctor s name patient s name and the date and start time of the test A Sharpie fine point erasable marking pen is recommended This information should also be noted in the patient s diary and the patient should then be instructed on how to fill out this diary Note Since the flash memory card is intended to be reused the patient label on...

Страница 10: ...onds Release the pushbutton and the LED will start blinking at a slower rate as described in Step e above h Close the top cover of the recorder place the recorder in the pouch and attach the recorder to the patient utilizing either the belt or the shoulder strap Caution the patient to avoid getting the recorder wet and to NOT pull on the electrodes i If there is a need to stop the recorder after t...

Страница 11: ...en as desired 6 1 Replacing the Clock Battery The digital clock uses one 1 5V silver oxide battery Duracell 392B or its equivalent The battery has approximately a two year life The battery is accessed under the top cover a Open the top cover Use a small jeweler s screwdriver inserted into the exposed center screw on the back of the clock module to gently lift the clock out of the endcap Figure 2 R...

Страница 12: ...of the recorder Braemar Inc 1285 Corporate Center Drive Suite 150 Eagan MN 55121 USA Phone 800 328 2719 Fax 651 286 8630 E mail service braemarinc com Web http www braemarinc com 7 1 Problem Description and Recommended Solution Problem Green LED is not blinking Solution Ensure that a Flash Memory card of at least 10 MB capacity has been properly inserted in the recorder Install a new set of AA alk...

Страница 13: ...the base of the lead to remove it from the patient input connector DL700 Holter Monitor 20 8 Service Items Part Number 1 HeartStor card 16 Mbytes 350 0147 01 2 HeartStor card 32 Mbytes 350 0147 02 3 HeartStor card 64 Mbytes 350 0147 03 4 Digital clock module 800 0364 01 5 Lead set patient 7 lead 200 1380 001 6 Lead set patient 5 lead 200 1380 002 7 ECG Test cable box 200 1588 001 8 Pouch with stra...

Страница 14: ...ed Channel 1 Place on the left Anterior Axillary line 6th rib Black Channel 2 Place on the left Manubrial border of the Sternum Brown Channel 2 Place at approximately 1 inch right of the Xiphoid Process on the rib Blue Channel 3 Place on the Center of the Manubrium Orange Channel 3 Place on the left Mid Clavicular line 6th rib Green Ground Place on the lower right rib margin over bone Channel Desi...

Страница 15: ...cement Chart White Channel 1 Place on the right Manubrial border of the Sternum Red Channel 1 Place on the left Anterior Axillary line 6th rib Black Channel 2 Place on the left Manubrial border of the Sternum Brown Channel 2 Place at approximately 1 inch right of the Xiphoid Process on the rib Green Ground Place on the lower right rib margin over bone Channel Designation Channel 1 White Red Channe...

Страница 16: ... 25 Figure 5 5 Lead Electrode Placement Braemar Inc Phone 800 328 2719 1285 Corporate Center Drive Suite 150 Fax 651 286 8630 Eagan MN 55121 USA E Mail Service BraemarInc com Copyright 2006 Braemar Inc All rights reserved ...