![BPS C-100 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/bps/c-100/c-100_user-manual_2798818009.webp)
5
HUMIDIFIER USAGE
The use of heated humidification or heated tubing will significantly
reduce the expected run time of the Battery Pack due to the amount of
power required for the humidifier heating plate.
To obtain optimal Battery Pack performance we recommend you forgo the
use of heated humidification. Using cool or “passover” humidification will
not reduce expected run times.
Please note that some manufacturers do not recommend the use of
humidification when an external battery is used to power their device or
they may recommend the use of a modified or pure sine wave power
inverter in conjunction with the battery.
Please check the manual for your
specific PAP device for disclaimers regarding humidification usage.
POWER BACKUP USAGE
Standard smart chargers keep the Battery Pack topped off and ready to go
as an emergency power backup during power outages.
If you know that a power outage is eminent or probable, you may connect
your device to the Battery Pack and plug the Battery Pack into a wall
outlet. In the event you lose power the Battery Pack will automatically
begin powering your device.
Do not consistently use the Battery Pack as a power backup option as
this will reduce its overall life. Only use the Battery Pack in this
manner when you know that a power outage is eminent.
I
mportant
N
otes
Содержание C-100
Страница 1: ...C 100 USER S GUIDE ...
Страница 27: ...1 877 445 5228 info batterypowersolutions net www batterypowersolutions net ...
Страница 28: ...CPAP BATTERY PACK USER S GUIDE January 2014 ...