5 Operation
MN031-443-S0-20110112-EN
TissueSeal / TissueSeal PLUS
21
Grasping, clamping and sealing tissue
WARNING
Risk of patient injury due to inadvertent activation of the
ligation instrument!
X
Never use the AUTOSTART function.
X
Do not switch on the HF current before the active
electrodes are in contact with the tissue to be
coagulated.
1. Place the jaws
A
on the operating site.
2. Place the tissue to be sealed between the electrodes of
the jaws
A
.
3. Close the jaws
A
to grasp the tissue.
ª
The tissue is grasped.
4. Use the two-position ratchet of the handle
B
to adjust
the pressure on the tissue to best match the amount of
grasped tissue.
ª
The tissue is clamped.
5. Using the foot switch of the HF device, activate the HF
current for coagulation:
•
A continuous acoustic signal sounds during
the entire sealing process to indicate that
power is being supplied.
•
An alternating acoustic signal indicates the
end of the sealing process.
6. Release the foot switch to stop the supply of power.
7. Using the handle
B
, open the jaws
A
.
ª
The tissue is coagulated.