4 Prearrangement
26
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
4.4.
Functional test
4.4.1. Auto test function
The HF device automatically performs cyclic testing during operation. If any faults
occur, see Section Detecting and correcting faults, page 73.
4.4.2. Functional testing
Perform the following functional test before putting the device into service:
The accessories must be designed for the specified maximum voltage.
1. Connect the neutral electrode and attach it securely to the patient’s arm.
The EASY neutral electrode indicator changes to green.
2. Remove the neutral electrode.
The indicator changes to red, acoustic signals sound.
The neutral electrode used for this test may not later be used for an
operation.
3. Connect a monopolar HF handpiece to a monopolar socket connector if there
is a green EASY indicator and use the manual switch and footswitch to
individually activate "Cut" and "Coag".
4. Check the settings on the display.
5. Now change to the bipolar output and connect bipolar forceps.
6. Select a mode with AUTOSTART, grasp moist gauze with the forceps, and
check the display.
7. Now change to a mode without AUTOSTART and use the footswitch to
activate the bipolar output. Check the settings and indicators in the bipolar
section.