11 Technical specifications
40
ARC 100 Operating Manual
900-100_IFU-V1.0_11587-S0-20141209-EN
Monitoring of the neutral electrode
EASY: Electrode Application System
Indicator for non-split or split electrode on the front
panel
Warning signal in the event of risk under the neutral
electrode
Visual, acoustic
Warning indicator on the front panel
Safety features
ISSys
:
I
ntegrated
S
afety
Sys
tem
Self-test
Continuous status indication on the front panel
Display of operating errors on the front panel
Display of system errors on the front panel
Service support
Service support via ISSys
Cooling
Convection
Duty factor
Intermittent
10 sec / 30 sec (on/off)
Characteristics
Max. CUT power
100 W (at 500 Ω)
Max. COAG power
100 W (at 100 Ω)
Output frequency
500 kHz
Sockets
1x monopolar/bipolar multifunctional
socket, monopolar with foot switch
and finger switch, bipolar with foot
switch
Connection for foot switch
1x
Scope of delivery
Incl. User manual, mains cable
Compatibility
Permitted combination
Foot switch (REF 901-012)
Conditions of operation, transport and
storage
Operation
Transport and
storage
Temperature
+10°C to +40°C
-10°C to +60°C
Relative humidity
30 to 75%, non-
condensing
10 to 85%, non-
condensing
Atmospheric pressure
700 to 1060 hPa
500 to 1060 hPa
Operating altitude (max.)
3600 m above sea level