Page 5
MN031-426-S2
Medical foot switch
10. Technical data
Standards:
DIN EN 60601-1:1990+A1+A2
IEC 60601-1:2007+A1+A2
DIN EN 60601-2-2:2007
IEC 60601-2-2:2010
DIN EN 60601-2-22:2005+A1:2007
IEC 60529
DIN EN 60529:1991+A1
Classification:
class IIb as per Medical Device Act / MDD 93/42/EEC
Housing and top (operating element):
unbreakable, self extinguishing thermoplast
Connection cable:
firmly attached and poured control line
Protection:
IP X8 as per IEC 60529
Switching element:
Reed contact
Microswitch
HALL sensor
(1S, 2S, 1W)
(1PW)
(HS 0-10 V, 0(4)-20 mA)
Voltage:
max. 25 V AC / 60 V DC
Supply voltage:
----
----
15 V DC < U
b
> 30 V DC
Current consumption:
max. 1 A
max. 5 A
max. 60 mA
Breaking capacity:
max. 100 VA
max. 1250 VA
----
Mechanical life:
> 1 Mio. switching cycles
Approval:
suitable for AP area
Temperature
Relative humidity
Air pressure
Storing conditions:
-
20 °C to +70 °C
10 - 100%
500 - 1060 hPa
Ambient
conditions:
-
10 °C to +60 °C
10 - 100%
800 - 1060 hPa