Bovie DERM 941 Скачать руководство пользователя страница 2

1

User’s Guide • DERM 941 / 942

USER’S GUIDE

Содержание DERM 941

Страница 1: ...DERM 941 DERM 942 High Frequency Desiccators User s Guide...

Страница 2: ...1 User s Guide DERM 941 942 USER S GUIDE...

Страница 3: ...8 Inspecting the Desiccator and Accessories 8 Setup Procedures 9 Performance Checks 11 Maintenance 11 Sterilization and Cleaning of the Accessories 11 Accessories 11 Technical Description 11 IEC Class...

Страница 4: ...tructions in these warnings and cautions It is also important that you read understand and follow the instructions for use in this user s guide Emergo Europe Molenstraat 15 2513 BH The Hague The Nethe...

Страница 5: ...l damage to tissue especially during use on small structures Use electrosurgery with caution in the presence of internal or external devices such as pacemakers or pulse generators Interference produce...

Страница 6: ...orate before the application if HF surgery There is a risk of pooling flammable solutions under the patient or in body depressions such as the umbilicus and in body cavities such as the vagina Any flu...

Страница 7: ...d be rated for at least the maximum peak output voltage of the H F generator set at the intended output control setting in the intended operating mode The output power selected should be as low as pos...

Страница 8: ...sthetic applied Intended User Profile Education Trained physician physician s assistant nurse nurse practitioner No maximum Knowledge Minimum Understands electrosurgery and electrosurgical techniques...

Страница 9: ...standard handpiece in the monopolar output as the footswitch allows you to activate the unit by either the handpiece or the footswitch Bipolar outputs are available for those physicians who prefer to...

Страница 10: ...lug the male end of the power cord into a grounded wall receptacle 4 The monopolar output for the handpiece is on the lower right front of the unit see figures 2 and 3 letter B The handpiece plug is d...

Страница 11: ...letter F When power level adjustment is being made by the handpiece an audible tone will sound to indicate that the power level has been changed Depressing and holding the up or down buttons will cau...

Страница 12: ...ied biomedical technician should conduct this testing to ensure that the unit is operating effectively and safely After the unit has passed the preliminary functional test it is ready for performance...

Страница 13: ...VAC Hospital grade power cord 10ft 3 048m 1 pcs For 110VAC models only 220VAC cord to be shipped only with special order units IP 55 239 User s Service Manual CD 1 pcs All models NOTICES A902 Handpiec...

Страница 14: ...is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Bovie DERM 941 942 can help prevent electromagnetic interference...

Страница 15: ...al or hospital environment If the user of the Bovie DERM 941 942 requires continued operation during power mains interruptions it is recommended that the Bovie DERM 941 942 be powered from an uninterr...

Страница 16: ...o AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey shou...

Страница 17: ...ation Error RESERVED Switch unit off and on again E9 Multiple Errors Switch unit off and on again Fault Code Description of Fault Recommended Action F1 Activation upon power up Check handpiece for act...

Страница 18: ...3 watts for power settings 1W 17 User s Guide DERM 941 942 Ambient temperature range 10 to 40 C Relative humidity 30 to 75 non condensing Atmospheric pressure 70kPa to 106kPa Warm up time If transport...

Страница 19: ...veforms as viewed on an oscilloscope Figure 11 illustrates output power delivered to rated load for all available modes at selected power settings Figure 5 Power setting versus voltage Vpeak Monopolar...

Страница 20: ...19 User s Guide DERM 941 942 Figure 11 Output Power VS Power Settings at Rated Load...

Страница 21: ...art Non ionizing radiation Neutral Electrode referenced to earth Volume control Manufacturer Mandatory Refer to instruction manual guide Conforms to European Union medical Directives 93 42 ECC and its...

Страница 22: ...U S Telephone 1 800 537 2790 Int l Telephone 1 727 384 2323 BovieMedical com sales boviemed com MC 55 239 001 Rev 0 yyyy mm dd...