Boston Scientific rezum Series Скачать руководство пользователя страница 4

Boston S

cientific (Master Brand DFU 

Template 8.2677in x 1

1.6929in 

A4, 9223851

9A), eDFU

, MB, R

ez

ū

m, en, 50849555-0

1A

Black (K) ∆E ≤5.0

Caution:

 If finger is released from Vapor Activation Button 

before the priming cycle is complete, vapor will automatically 

stop, and the priming steps will have to be repeated.

F.  If the Vapor Activation Button is released before the 

end of the priming cycle, repeat the priming cycle 

(steps A to E).

G.  If priming cycle is not successfully completed, repeat 

steps a to e or replace Delivery Device.

Perform the Pre-Treatment Vapor Cycle

1.  Activate idle feature by running a pre-treatment vapor 

cycle. Idle feature heats coil to keep water in a ready 

state so vapor delivery is immediate. If this step is 

not completed, condensation may build up between 

treatments, which may lead to insufficient treatment.

2.  Pull in Flush Activation Button (1) Needle Deployment 

Button (2) and then Vapor Activation (3) (Figure 12).

3.  During pre-treatment vapor cycle, observe flush exiting tip.
4.  When pre-treatment vapor cycle is complete, release the 

Vapor Activation Button and retract the needle by pushing 

upward on the Needle Retraction Button.

Note:

 Pre-treatment vapor cycle must be completed prior to 

inserting Delivery Device into the patient.

Perform the Rez

ū

m™ Vapor Treatment

1.  Confirm the Generator display is showing the Therapy Screen.
2.  Coat the shaft of the Delivery Device with water-soluble 

lubricating or anesthetizing gel.

3.  Attach light cord and video camera to the scope lens.
4.  Using finger, activate the saline flush by applying gentle 

pressure to the Flush Activation Button.

5.  Carefully insert the Delivery Device into the urethra 

through the meatus.

Warning:

 Excessive pressure while using Flush Activation 

Button may cause unintended deployment of the needle.

Warning:

 Ensure needle is fully retracted by viewing needle 

position through scope lens. If needle is not retracted prior to 

insertion of the Delivery Device, damage to urethra may occur.

Warning:

 No modification of this equipment is allowed. Do not 

attempt to service or maintain the generator while in use with 

a patient.

6.  While examining prostatic urethra, locate the apex of the 

prostate and the bladder. A TRUS and/or cystoscopy prior 

to the procedure can help obtain prostate measurements 

to determine the appropriate number of treatments.

7.  Estimate the prostatic treatment length (i.e. from bladder 

neck to verumontanum). This length is considered the 

vapor treatment zone (Figure 13).

Bladder Neck

Verumontanum

Figure 13. Prostatic Treatment Length. 

8.  Based on the length of the vapor treatment zone, determine 

the number of treatments per lobe (Table 2). A treatment 

consists of a single 9-second delivery of vapor.

Table 2. Guidelines for determining the number of treatments 

(lateral lobe).

Distance from Bladder Neck 

to Veru

Estimated Number of 

Treatments per Lobe

< 2.0 cm

1-2

2.0 – 3.0 cm

2-3

> 3.0 cm

3-4

9.  If a median lobe is present and judged to be in need of 

treatment, deliver one treatment if median lobe is < 2 cm and 

two or more treatments if median lobe is > 2 cm. If central zone 

hyperplasia contributes to an elevated bladder neck with a 

prostatic urethral ≥ 35 degrees, as evidenced by sagittal TRUS, 

deliver one treatment for an enlarged central zone < 2 cm and 

two treatments for an enlarged central zone > 2 cm.

Caution:

 Treatments in excess of those recommended in the guidelines 

may lead to prolonged irritative symptoms and/or catheterization.

Note:

 A maximum number of 15 full treatments can be delivered 

with each Delivery Device.

Figure 14. Illustrative example of 6 Vapor Treatments. 

Warning:

 Proper placement of the needle is essential. Do not 

direct the needle downward toward the rectum.

10. Start the procedure by positioning the tip of the device just 

inside the bladder. Rotate the Delivery Device 90 degrees 

(horizontal) and bring device shaft just off floor of urethra.

11. While maintaining the 90 degree rotation, pull Delivery Device 

back into the urethra and position 1 cm back from the bladder 

neck. If treatment occurs within 1 cm of the bladder neck, 

short-term irritative symptoms may be experienced by the 

patient. Place the distal tip of the Delivery Device shaft against 

the lateral urethral wall.

Note:

 Optimal placement for the vapor treatment is in the crest of 

the lateral lobe. Ensure the shaft of the device is not close to the 

anterior surface, as this may lead to a sub-optimal treatment.

Note:

 On occasion, patient prostatic anatomy may restrict the 

Delivery Device tip from reaching the bladder neck. This may be 

due to an elevated bladder neck from central zone hyperplasia 

or a median lobe. On these occasions, do not force the device 

through tissue. Ensure the Delivery Device tip is proximal to the 

verumontanum and treat the bulk of the lateral lobe proximal to the 

verumontanum. Advance the Delivery Device in 1 cm increments 

toward the bladder neck to deliver subsequent vapor treatments. 

This may relax the tissue to allow the Delivery Device to reach the 

bladder neck. If the Delivery Device still cannot reach the bladder 

neck, treat the area that is proximal to the verumontanum.

12. Stabilize the Delivery Device before deploying the needle and 

remain completely still throughout the treatment.

13. While holding the Flush Activation Button, continue to pull in 

the Needle Deployment Button until the needle is deployed.

14. Visually verify the needle is fully inserted into the prostate by 

inspecting to see that the black depth marker just proximal to 

the emitter holes is not visible (no black should be seen).

Warning:

 Do not start treatment if the black depth marker on the 

needle is still visible after needle deployment. If the marker is still 

visible, push the needle deeper into the prostate until no black 

is visible through the lens. If unable to position correctly retract 

needle. Reposition Delivery Device approximately 1 cm from the 

partially treated site and repeat needle deployment steps.

15. Using finger, pull in Vapor Activation Button and hold to 

activate the vapor until treatment cycle is complete.

Caution:

 Once needle is deployed, hold the Delivery Device 

steady. Movement of the Delivery Device may stretch tissue and 

cause vapor to leak into the urethra, causing urethral irritation. 

Extreme movement may also cause pressure on the needle 

resulting in difficulty with needle retraction. Needle must be 

returned to the original insertion position to facilitate retraction.

Note:

 When the vapor treatment begins, the Rez

ū

m System 

automatically tracks the time until the programmed treatment 

is complete and then automatically shuts off the vapor. 

Vapor can be stopped prior to treatment completion if Vapor 

Activation Button is released.

Caution:

 Do not release Vapor Activation Button during vapor 

treatment cycle. If Vapor Activation Button is released before 

the cycle is complete, vapor release will automatically stop, 

which may lead to partial and incomplete treatment.

16. The display screen will show each individual treatment time 

and count the number of full treatments that were completed.

17. Release Vapor Activation button and push upward on the 

Needle Retraction Button to retract the needle.

Warning:

 Ensure needle is fully retracted by viewing needle 

position through scope lens. If needle is not retracted prior to 

repositioning Delivery Device, damage to urethra may occur.

18. Reposition the Delivery Device for the next treatment by 

moving the device tip approximately 1 cm distal to the 

previous needle placement. The objective is to create 

contiguous, overlapping lesions, 1 cm apart, and running 

parallel to the prostatic urethra.

19. Maintain device rotation at 90 degrees between treatments 

to avoid losing sight of previous treatment location.

20. Follow the natural slope of the urethra to avoid being too 

close to the ceiling, i.e. too anterior. Center the needle 

between the floor and ceiling of the urethra and target the 

bulk of the adenoma directly if it is not centered.

21. Complete steps 10-20 until all treatments in the first lateral 

lobe are complete. The final treatment location within each 

lobe should be on the proximal side of the verumontanum.

Warning:

 Prior to each treatment, know where the 

verumontanum is in relation to the tip of the shaft. All 

treatments should be placed proximal to the verumontanum.

22. Return Delivery Device to the start position at the bladder 

neck for treatments in the contralateral lobe. Rotate the 

Delivery Device 90 degrees to enable needle insertion at 

desired location on opposite lobe.

23. Repeat steps 10 through 20 until second lobe is fully treated.
24. For intravesical prostatic protrusions of either the lateral 

or median lobes, position Delivery Device 1 cm from the 

proximal edge of the protrusion and deliver the vapor 

treatment with the needle positioned approximately 45 

degrees toward the midline. One treatment for a small 

median lobe (< 2 cm) and two or more treatments for a 

larger median lobe (> 2 cm). For an enlarged central zone, 

deliver treatments 1 cm from the bladder neck with the 

needle positioned at 45 degrees toward the midline of the 

tissue. Do not treat on the floor of the urethra within at 

least 1 cm of the verumontanum.

Caution:

 Care should be taken during procedure to monitor 

remaining saline level. If saline source is empty, patient could 

experience urethral discomfort due to no flush flow.

25. With lens in place, visually inspect the urethra and bladder 

at the end of the treatment and withdraw the Delivery 

Device from the urethra.

26. To conclude procedure, select Remove Device on 

Generator screen and follow instructions.

Содержание rezum Series

Страница 1: ...e exterior of the Sterile Water Vial is not sterile and should not be placed in the sterile field Positioning Saline Flush Line in Saline Pump Reference indicators on Generator to ensure Saline Flush Line is positioned in the correct direction If Saline Flush Line is placed in a backwards direction within the Saline Pump saline will not flow during procedure Remaining Saline Level in Bag Care shou...

Страница 2: ...tray Topical antiseptic e g Betadine Patient drape Disposable underpads e g Chux Gauze squares Lidocaine gel anesthetic or water soluble lubricating gel Saline supply at room temperature 1 L 2 L 3 L 4 L 5 L or 500 ml IV pole for Saline supply 4 mm 30 degree 30 cm Storz or Richard Wolf rigid cystoscope lens Light source and cord Video camera and display recorder optional Drain bucket Hemostat Prepa...

Страница 3: ...sh Line is placed in a backwards direction within the Saline Pump saline will not flow during procedure 5 Close Saline Pump door prior to attaching Saline Flush Line tip to the Saline bag Note If Saline Flush Line tip is attached to Saline bag prior to placing Saline Flush Line in the Saline Pump and closing the Saline Pump door saline may leak 6 Remove cap from tip of Saline Flush Line and attach...

Страница 4: ...ry Device tip is proximal to the verumontanum and treat the bulk of the lateral lobe proximal to the verumontanum Advance the Delivery Device in 1 cm increments toward the bladder neck to deliver subsequent vapor treatments This may relax the tissue to allow the Delivery Device to reach the bladder neck If the Delivery Device still cannot reach the bladder neck treat the area that is proximal to t...

Страница 5: ...he shaft to manually retract the proximal end of the needle into the shaft tip Fig 17 4 While maintaining the needle tip within the shaft remove Delivery Device from patient 5 If treatment is incomplete re start procedure with new Delivery Device and complete procedure 6 Report all incidences of manual needle retraction to Boston Scientific Customer Service Clean device and return to Boston Scient...

Страница 6: ...Procedure or Device Related AEs Severity Resolved AEs Mild Moderate Severe Dysuria 17 14 4 0 97 Hematuria Gross 12 11 1 0 100 Hematospermia 6 6 1 0 100 Urinary Frequency 6 5 1 0 82 Decrease in Ejaculatory Volume 5 4 1 0 33 UTI Suspected 5 4 1 0 100 Urinary Retention 5 1 4 1 100 Urinary Urgency 5 3 2 0 78 Anejaculation 3 2 1 0 0 Terminal Dribbling 3 2 1 0 60 UTI Culture Proven 3 1 2 0 100 Epididymi...

Страница 7: ... M006D2201 003 G2200 003 GTIN 08714729992547 GTIN 0855357006003 Responsible Party U S Contact Information Boston Scientific Corporation 300 Boston Scientific Way Marlborough MA 01752 508 382 9555 www bostonscientific com FCC Compliance Statement This device complies with Part 18 of the FCC Rules For further information see FCC web site for a complete description of all requirements WARRANTY Boston...

Страница 8: ... REP Legal Manufacturer Boston Scientific Corporation 300 Boston Scientific Way Marlborough MA 01752 USA USA Customer Service 888 272 1001 Recyclable Package Australian Sponsor Address Boston Scientific Australia Pty Ltd PO Box 332 BOTANY NSW 1455 Australia Free Phone 1800 676 133 Free Fax 1800 836 666 AUS Do not use if package is damaged Argentina Local Contact ARG Para obtener información de con...

Отзывы: