To Access the Pulse Width and Rate Options (if enabled through Bionic Navigator):
1. Press and hold the
"
button on the Remote Control for three seconds until the Pulse Width
screen appears. This is indicated by “Width” in the bottom left corner of the screen.
2. Area 1 will initially be highlighted. Press the
"
button to cycle to the Area you wish to adjust.
3. When the desired Area is highlighted, press
p
or
q
to increase or decrease the Pulse
Width.
4. To continue to Rate adjustment, press and hold the
"
button for three seconds from any
Pulse Width Area screen. The Rate screen is indicated by “Rate” in the bottom left corner of
the screen.
5. Area 1 will initially be highlighted. Press the
"
button to cycle to the Area you wish to adjust.
6. When the desired area screen is highlighted, press
p
or
q
to increase or decrease the
Rate.
7. To return to the main Level screen, press and hold the
"
button for three seconds until the
Level screen appears.
Device Linking
A Remote Control can communicate with only one Stimulator at a time. This prevents the
Remote Control from accidentally controlling an unintended device. In the linking process, the
Remote Control will identify, by telemetry, the intended Stimulator for communication.
1. Press and hold the
P
button to unlock the Remote Control and initiate communication
between the Remote Control and the Stimulator.